Precise Digital Accelerator recallZ-1088-2015 · Elekta, Inc. · initiated Jan 7, 2015
Class II Terminated All recalls posted in February 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Elekta, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
FDA root cause category: Other
What the company is doing
Elekta sent an Urgent Important Field Safety Notice on January 7, 2015, to all affected customers. Elekta recommended that customers replace the M20 x 50 MM bolts marked 'OF' on the gantry counterbalance frame immediately. The digital linear accelerator must not be used until these 'OF' bolts are replaced. Elekta recommended that customers put the machine in a parked position at 0, and attach the gantry locking straps. A Modification kit will be available as soon as possible which contains the correct bolts. Customers were instructed to complete the Important Field Safety Notice Acknowledgement form and return to their local Elekta Office or Representative. Customers with questions were instructed to contact their local Elekta representative. For questions regarding this recall call 404-993-5886.
Recall details
- Recall number
- Z-1088-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 7, 2015
- Posted by the FDA
- Feb 10, 2015
- Recalling firm
- Elekta, Inc.
- Quantity
- 46
- Product code
- IYE
- FDA submissions
- K051932
- Distribution
- Worldwide Distribution - USA including AZ, CA, FL, KS, LA, MN, MO, NY, NC and Internationally to Austria, Bulgaria, Canada, China, Germany, Hungary, Italy, Malaysia, Malta, Netherlands, Norway, Russia, Spain, Sweden, Taiwan and Ukraine.
Product description
Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.
Code information (models, lots, serial numbers)
153602, 153538, 153561, 153569, 153532, 153556, 153566, 153563, 153564, 153589, 153530, 153531, 153562, 153584, 153615, 153534, 153534, 153535, 153563, 153542, 153542, 153548, 153548, 153558, 153578, 153595, 153699, 153630, 153728, 153680, 153712. 153717, 153723, 153734, 153702, 153715, 153684, 153710, 153577, 153725, 153726, 153691, 153696, 153713, 153724, 153727
Other devices under the same FDA submission
1 device cleared under K051932 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- NA 1525717Elekta Limited · Accelerator, linear, medical1 recall3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.