Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004… recallZ-1324-2015 · Titan Spine, LLC · initiated Feb 20, 2015

Class II Terminated All recalls posted in March 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Titan Spine, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The TCS Straight Hexlobular #6 Drivers (5210-1004) tip sheared off during use in procedures that involved patients with dense bones. As a result of this malfunction, Titan Spine, LLC is removing all of the TCS Straight Hex lobular #6 Drivers (5210-1004) Lot# G150101 from the field.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Titan Spine sent an "Urgent Medical Device Recall" letter via Fed Ex dated February 20, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A replacement TCS Straight Hex lobular# 6 Drivers (5210-1004) will be shipped out to each Titan Spine Area Vice President and Sales Rep. within 4-6 weeks. For further questions please call (262) 242-7801.

Recall details

Recall number
Z-1324-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 20, 2015
Posted by the FDA
Mar 26, 2015
Recalling firm
Titan Spine, LLC
Quantity
26
Product code
OVE
FDA submissions
K142940
Distribution
US Distribution to the states of : AZ, CA, CT, FL, GA, ID, IL, KY, MD, MO, NV, NY, OH, OK, OR, TX, UT and WI.

Product description

Endo TCS Driver Straight Hexalobe #6 P/N 5210-1004, Non-Sterile, Manufactured by Titan Spine, LLC. Endo TCS driver is one of the instruments within the surgical tray for use with the Interbody Fusion Device(IBD).

Code information (models, lots, serial numbers)

G150101

Other devices under the same FDA submission

5 devices cleared under K142940 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.