Variable Angle Locking Peg, D:1.8mm x L:14.0mm recallZ-1314-2015 · Flower Orthopedics Corporation · initiated Feb 16, 2015
Class II Terminated All recalls posted in March 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Flower Orthopedics Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Locking pegs were not locking during distal radius procedures. There has been one (1) report of revision surgery scheduled to remove a loosened peg.
FDA root cause category: Device Design
What the company is doing
On February 16, 2015, Flower Orthopedics generated and distributed a field memo/advisory notice (FOCMB-15-02) to all distributors with locking peg inventory on February 6, 2015. The field memo provided technique guidance to end users. All 12 consignees with remaining inventory (ie.not reported as used) were contacted directly via phone and all have agreed to ship the locking pegs back to Flower Orthopedics. Flower Orthopedics determined that a generic statement to all customers was not necessary for the following reasons detailed above and summarized again here: product was distributed only through direct channels, the health hazard evaluation has determined there is a very low risk to health, and all consignees have been directly contacted already and product inventory is being returned.
Recall details
- Recall number
- Z-1314-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 16, 2015
- Posted by the FDA
- Mar 24, 2015
- Recalling firm
- Flower Orthopedics Corporation
- Quantity
- 901
- Product code
- HWC
- FDA submissions
- K123562, K131657
- Distribution
- Distributed in the states of VA, TX, IL, OK, PA, CA, NY, KY, and FL.
Product description
Variable Angle Locking Peg, D:1.8mm x L:14.0mm; Variable Angle Locking Peg, D:1.8mm x L:16.0mm, Variable Angle Locking Peg, D:1.8mm x L:18.0mm; Variable Angle Locking Peg, D:1.8mm x L:20.0mm; Variable Angle Locking Peg, D:1.8mm x L:22.0mm, Variable Angle Locking Peg, D:1.8mm x L:24.0mm; Variable Angle Locking Peg, D:1.8mm x L:26.0mm. The locking pegs are intended to be used in conjunction with the Flower Orthopedics Distal Radius Plate for osteosynthesis of the radius bone.
Code information (models, lots, serial numbers)
All lots of Catalog Numbers: FLP 114, FLP 116, FLP 118, FLP 120, FLP 122, FLP 124, FLP 126
Other devices under the same FDA submission
14 devices cleared under K123562, K131657 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Flower E-Kit, Advanced EWK 201Flower Orthopedics Corporation · Pin, Fixation, Smooth1 recall0 reports
- Distal Radius Plate, Standard, Left, 3 Holes FDR 011Flower Orthopedics Corporation · Plate, Fixation, Bone2 reports
- Variable Angle Non-Locking Screw FBS 355Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
- Variable Angle Locking Screw FRP 316Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
- Variable Angle Locking Screw FRP 318Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
- Variable Angle Non-Locking Screw FBS 508Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
- Distal Radius Plate, Dorsal (R), 9 Holes FRP 432Flower Orthopedics Corporation · Plate, Fixation, Bone1 report
- Variable Angle Locking Screw FLP 322Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
- Variable Angle Locking Screw FLP 326Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
- MTP Plate, 5 Holes, Left MFP 012Flower Orthopedics Corporation · Plate, Fixation, Bone1 report
- Variable Angle Locking Peg FDR 118Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
- Variable Angle Locking Peg FDR 120Flower Orthopedics Corporation · Screw, Fixation, Bone1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.