ADVIA Chemistry XPT System, Siemens Material Number… recallZ-1399-2015 · Siemens Healthcare Diagnostics, Inc. · initiated Feb 11, 2015
Class II Terminated All recalls posted in April 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Siemens Healthcare Diagnostics has confirmed an issue with ADVIA Chemistry XPT software version 1.0.2. The system may cause samples to remain in an Inprocess state. Test results on a sample that is held Inprocess will not transmit to the LIS. Manual intervention is necessary to complete the processing of the samples that are held Inprocess.
FDA root cause category: Software design
What the company is doing
Siemens Healthcare Diagnostics sent Urgent Medical Device Correction Letters (dated 2/11/2015) via FedEx to the domestic consignees informing users of the software issue with ADVIA Chemistry XPT Software Version 1.0.2. The letters included work instructions to handle the bug. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within 30 days. For questions about in process samples, the event message 02 981 8728, or need assistance with mitigation solutions, please contact your Siemens Customer Care Center or your local Siemens Technical Support Representative. Urgent Field Safety Notice Letters were sent to the foreign consignees (distribution is determined at the country level). Field service personnel were notified of the Urgent Medical Device Correction Letter.
Recall details
- Recall number
- Z-1399-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 11, 2015
- Posted by the FDA
- Apr 7, 2015
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 16 units: 3 Domestic, 13 Foreign
- Product code
- JGS
- FDA submissions
- K990346
- Distribution
- Distributed in the states of AZ, CA, and WA, and the countries of Germany, Italy, Spain, and UK.
Product description
ADVIA Chemistry XPT System, Siemens Material Number 10723034, Software Version 1.0.2, Siemens Material Number 11219493, IVD. The ADVIA XPT Chemistry System is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes at a throughput rate of both 1800 photometric tests per hour and 600 electrolyte (ISE) tests per hour.
Code information (models, lots, serial numbers)
Software Version 1.0.2; Lot Numbers: CA1275000080008, CA1275000090009, CA1275000060006, CA1275000050005, CA1275000070007, CA1275000130013, CA1275000100010, CA1275000120012, CA1275000170017, CA1275000260026, CA1275000270027, CA1275000280028, CA1275000240024, CA1275000200020, CA1275000220022, CA1275000410041
Other devices under the same FDA submission
2 devices cleared under K990346 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ADVIA® Chemistry XPT System 10723034Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE1 recall174 reports
- ADVIA® 1800 Analyzer 10315183Siemens Healthcare Diagnostics Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE150 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.