Quadrox-iD Adult Diffusion Membrane Oxygenator recallZ-1689-2015 · Maquet Cardiovascular Us Sales, Llc · initiated Mar 5, 2015
Class II Terminated All recalls posted in May 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A potential issue was detected that could result in blood leakage from the gas outlet port in all lots of Quadrox-iD Adult Diffusion Membrane Oxygenator with Bioline and Softline Coatings, Custom Tubing Packs and Diffusion Membrane HLS Set Advanced and HIT Set Advanced Models.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Maquet sent a "Field Safety Notice" accompanied by a Response Form to their affected customers on 3/5/2015 via FedEx. The Field Safety Notice identified the affected products/Part numbers and actions to be taken by the user. The customers were asked to complete the Field Safety Notice Response Form to acknowledge that they have received the notice. The completed notice can be faxed to 1-973-396-3607 or sent via email to [email protected]. For any questions, customers can contact their Maquet Sales Representative or Maquet Customer Service at 1-888-627-8383 (press option 2, followed by option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. EST.
Recall details
- Recall number
- Z-1689-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 5, 2015
- Posted by the FDA
- May 28, 2015
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Quantity
- 10,199 units
- Product code
- DTZ
- FDA submissions
- K132829
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Great Britain, Greece, Hungary, Iceland, Ireland, Israel, Italy, Japan, South Korea, Latvia, Lithuania, Luxembourg, Malta, Mexico, Netherlands, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia Montenegro, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey and Uruguay
Product description
Quadrox-iD Adult Diffusion Membrane Oxygenator Product Usage: The Quadrox-iD Adult Diffusion Membrane Oxygenator is intended for use in an extracorporeal circulation during cardiopulmonary bypass in cardiac surgery. Within the specified flow rate range, the device oxygenates the blood, removes carbon dioxide from the blood and regulates the blood temperature. The oxygenator with diffusion membrane cannot be used for the administration of volatile anesthetic gases because the gas molecules are unable to pass through the diffusion membrane. Responsibility for deciding whether to use an oxygenator rests solely with the attending physician.
Code information (models, lots, serial numbers)
Part numbers BO-HMOD 70000, 70105.3815, BEQ-HMOD 70000, 70105,3824
Other devices under the same FDA submission
2 devices cleared under K132829 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- QUADROX-i Pediatric, diffusion, BIOLINEMaquet Cardiopulmonary AG · HEAT-EXCHANGER, CARDIOPULMONARY BYPASS2 recalls0 reports
- Quadrox-iD BEQ-HMOD30000-USAMaquet Cardiopulmonary AG · Heat-Exchanger, Cardiopulmonary Bypass1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 70681).
- Z-1690-2015 · TerminatedHLS/HIT Set Advanced with Bioline and/or Softline Coating Console Product Usage: HLS/HIT Set Advanced: The CARDIO HELP System in configuration with the HLS/HIT Set Advanced is in…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page