MicroScan Microbiology Systems: Dried Gram Negative… recallZ-1531-2015 · Beckman Coulter, Inc. · initiated Aug 8, 2014
Class II Terminated All recalls posted in April 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Technical Support Bulletin issued in February 2013 did not mention performance limitation for Proteus mirabilis and Imp when using the CLSIM100-S20 revised interpretive criteria.
FDA root cause category: Labeling False and Misleading
What the company is doing
The firm, Siemens, sent an "Urgent Medical Device Correction" letter dated August 7, 2014 to affected customers by trackable method. The letter describe product, problem and actions to be taken. The customers were instructed to confirm receipt of the letter; complete the attached "Field Correction Effectiveness Check" form via fax to Siemens Healthcare Diagnostics at 302-631-8467;TSB 190 has been removed from use and archived from the Siemens Document Library, if you have a hardcopy, discard it appropriately;and follow the listed in instructions and retain and forward this letter to those who may have received this product. If you have any questions, please contact your Siemens Customer Care Center; Quality Systems & Compliance Director at 916-374-3031 or email: [email protected] or your local Siemens technical support representative.
Recall details
- Recall number
- Z-1531-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Aug 8, 2014
- Posted by the FDA
- Apr 27, 2015
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- Approx ~(216240 boxes U.S. and 84249 boxes OUS) total
- Product code
- JWY
- FDA submissions
- K870074
- Distribution
- Worldwide Distribution: US (Nationwide) and to countries of: Argentina, Bahamas, Brazil, Canada, China, Ecuador, Egypt, Germany, Lithuania, Mexico, Paraguay, Peru Poland, Portugal, Romania, Russian Federation, Singapore, South Africa, Spain, Taiwan, Thailand, Trinidad and Tobago, United Arab Emirates, Uruguay, and Venezuela.
Product description
MicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/Urine Combo 55, Catalog B1017-409 Microbiology: For use with MicroScan Dried Gram Negative MIC/Combo Panels and Dried Gram Negative Breakpoint Combo Panels. MicroScan panels are designed for use in determining antimicrobial agent susceptibility and/or identification to the species level of aerobic and facultatively anaerobic gram-negative bacilli.
Code information (models, lots, serial numbers)
Catalog number: B1017-409; Siemens Material Number (SMN): 10371533.
Other devices under the same FDA submission
1 device cleared under K870074 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- MicroScan B1017-216Beckman Coulter, Inc. · Susceptibility Test Cards, Antimicrobial1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 70893).
- Z-1528-2015 · TerminatedMicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg Combo 50, Catalog B1017-…
- Z-1529-2015 · TerminatedMicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/BP Combo 47, Catalog B10…
- Z-1530-2015 · TerminatedMicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/Urine Combo 51, Catalog …
- Z-1532-2015 · TerminatedMicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg/Urine Combo 61, Catalog…
- Z-1533-2015 · TerminatedMicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: Neg MIC 38, Catalog B1017-41…
- Z-1534-2015 · TerminatedMicroScan Microbiology Systems: Dried Gram Negative MIC/Combo Panels Containing Lower Imipenem Dilutions for Reporting CLSI M100-S20 Breakpoints Panel: ESL plus, Catalog B1027-101 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page