Dimension EXL with LM, Dimension EXL 200, Dimension RxL… recallZ-1615-2015 · Siemens Healthcare Diagnostics, Inc. · initiated Apr 2, 2015

Class II Terminated All recalls posted in May 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain Dimension Integrated Chemistry Systems Instrument Power Configurations may have been installed with a non-standard instrument power plug.The use of a non-standard plug configuration as described may lead to a potential electrical hazard. The likelihood of potential clinical impact or injury due to this issue is extremely low.

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens sent an Urgent Field Safety Notice, dated April 2015, was issued for distribution to affected customers in Japan to notify them of the issue. A Siemens Service Representative will visit customer sites to inspect the Dimension instrument power plug and wall outlet configuration. If the instrument is configured with a non-standard 15 Amp plug/outlet or a 20 Amp twist lock plug, Siemens will correct the issue identified, ensuring that the system is installed according to the instrument requirements.

Recall details

Recall number
Z-1615-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 2, 2015
Posted by the FDA
May 14, 2015
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
17801
Product code
JJE
FDA submissions
K073604, K112999, K130276
Distribution
Distributed in the country of Japan.

Product description

Dimension EXL with LM, Dimension EXL 200, Dimension RxL, Dimension RxL HM, Dimension RxL Max, Dimension RxL Max HM, Dimension Xpand, Dimension Xpand HM, Dimension Xpand Plus, Dimension Xpand Plus HM. The Dimension EXL clinical chemistry system is a discrete, random-access, microprocessor-controlled, integrated instrument/chemistry system that measures a variety of analytes, including enzyme activities, in body fluids for in vitro diagnostic use.

Code information (models, lots, serial numbers)

Model numbers: D045912, D193440, D140066;  Material Numbers: 10472176, 10636928, 10465412, 10461038, 10444828, 10444829, 10461849, 10461043, 10444837, 10444838

Other devices under the same FDA submission

2 devices cleared under K073604, K112999, K130276 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.