FLOW-i Anesthesia System recallZ-1706-2015 · Maquet Cardiovascular Us Sales, Llc · initiated Apr 2, 2015
Class I Terminated All recalls posted in June 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Maquet Cardiovascular Us Sales, Llc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
In some Flow-i Anesthesia Systems, patient cassette can get dislodged which could result in gas leakage within the system internal circuit potentially leading to a stop in ventilation.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm, Maquet Inc., sent an "Urgent Medical Device Field Action Field Correction for FLOW-i Anesthesia Systems" letter dated 4/2/2015 via FedEx. The letter describes the product, problem and actions to be taken. The customers were instructed to read the letter and inform staff, who may use the FLOW-i Anesthesia System of the patient cassette locking device issue. Customers are to obtain the serial number of their system; verify the serial number obtained against the serial number listed in the box in the Appendix. A Maquet Service Representative will be contacting you to arrange a convenient time to replace your FLOW-i Anesthesia System patient cassette locking device. If you have any questions regarding this field action, please contact your Maquet Service Representative or Maquet Technical Support at 1-888-627-8383 (Press option 3, followed by option 1 and then option 1 again), Monday through Friday, between the hours of 8:00 a.m. and 5:00 p.m. EST.
Recall details
- Recall number
- Z-1706-2015
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Apr 2, 2015
- Posted by the FDA
- Jun 8, 2015
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Quantity
- 1641 [80 units (US) 1561 units (OUS)]
- Product code
- BSZ
- FDA submissions
- K133958
- Distribution
- Class I Recall - Worldwide distribution -- US, including the states of AL, AZ, DC, FL, MI, MO, NJ, NY, OR, PA, TX, and WV; and, countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Lebanon, Libya, Mexico, Mongolia, Mozambique, Netherlands, Norway Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom, Unites States, and Venezuela.
Product description
FLOW-i Anesthesia System The FLOW-i Anesthesia System is intended for use administering inhalation anesthesia on neonatal to adult patient populations.
Code information (models, lots, serial numbers)
Article number 66 77 200 66 77 300 66 77 400 Serial Numbers 1058 1646 2034 2723 1059 1647 2095 2724 1060 1648 2186 2729 1172 1650 2554 2731 1201 1651 2555 2732 1202 1652 2556 2733 1204 1653 2557 2735 1224 1654 2558 2736 1225 1655 2559 2737 1226 1656 2560 2738 1277 1657 2561 2739 1280 1786 2562 2740 1281 1787 2613 2741 1282 1860 2714 2742 1545 1861 2716 2743 1546 2021 2719 2744 1643 2022 2720 2745 1644 2023 2721 2746 1645 2024 2722 2747
Other devices under the same FDA submission
2 devices cleared under K133958 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Vaporizer Sevoflur. Maquet filling 6682282Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall0 reports
- Vaporizer Sevoflurane, QUIK FIL 6682285Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page