Ultraview SL Command Modules, Model 91496, with the Masimo… recallZ-1577-2015 · Spacelabs Healthcare Inc · initiated Apr 21, 2015
Class II Terminated All recalls posted in May 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Modules resetting. This failure mode happens when the Spacelabs Command Module is configured with the Masimo SpO2 option (-M), the Spacelabs Respiration option (-R) and the Masimo SpO2 PCBA, PN: 010-1136-02. With these two options and the Masimo SpO2 PCBA present, the module may experience random resets.
FDA root cause category: Software design
What the company is doing
Spacelabs expanded the recall to include additional 29 units and notified one additional new consignee on June 4, 2015. ******************************************************************************************* The firm, Spacelabs Healthcare, sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated 27 April 2015, via priority mail to US consignees and email to foreign consignees. The letter described the product, problem and actions to be taken. The consignees were instructed to circulate this notification to all persons concerned by the information it contains; add a copy to the appropriated system user manuals until such time as corrections have been implemented by Spacelabs, and weigh the benefits versus the risks when deciding whether or not to continue to use your modules until they can be upgraded. Only modules shipped or repaired after March 8, 2014 are affected by this recall. Affected command modules will be upgraded with a new version of software by Spacelabs at no cost. Customer with questions about this corrective action, please contact Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical Support.
Recall details
- Recall number
- Z-1577-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Apr 21, 2015
- Posted by the FDA
- May 4, 2015
- Recalling firm
- Spacelabs Healthcare Inc
- Quantity
- 1973 units total (1345 modules in the US and 604modules international and 3 PCBAs in US and 21 PCBAs international)
- Product code
- DSI
- FDA submissions
- K103142
- Distribution
- Worldwide Distribution:US (nationwide) including states of: AL, AR, AZ, CA, FL, GA, IA, ID, IL, LA, ME, MI, MN, MO, NE, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WA, WY, and Puerto Rico and countries of: Argentina, Australia, Bangladesh, Canada, Chile, France, Great Britain, India, Kuwait, Panama, Poland, Qatar, Saudi Arabia, and Turkey. .
Product description
Ultraview SL Command Modules, Model 91496, with the Masimo SpO2 PCBA, PN: 010-1636-02. The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant populations in any type of clinical environment other than home use.
Code information (models, lots, serial numbers)
Show the full code information (26k characters)
FIRM EXPANDED RECALL ON JUNE 4, 2015 TO INCLUDE ADDITIONAL 29 SERIAL NUMBERS IN THE US: 1496-005416, 1496-005717, 1496-005727, 1496-005734, 1496-005735, 1496-005737, 1496-005739, 1496-005744, 1496-005745, 1496-005746, 1496-005760, 1496-008609, 1496-008614, 1496-008918, 1496-009096, 1496-009666, 1496-010906, 1496-014637, 1496-022087, 1496-023491, 1496-034466, 1496-036281, 1496-044908, 1496-103794, 1496-109373, 1496-109960, 1496-120322, 1496-122842, and 1496-127070. ******************************************************************************************* PCPA PN: 010-1636-02 **************SERIAL NUMBERS OF MODULES IN THE US****************** 1496-000595, 1496-005414, 1496-005702, 1496-005703, 1496-005704, 1496-005708, 1496-005710, 1496-005713, 1496-005714, 1496-005743, 1496-005748, 1496-005759, 1496-005786, 1496-006113, 1496-008032, 1496-008591, 1496-008595, 1496-008598, 1496-008599, 1496-008601, 1496-008602, 1496-008603, 1496-008605, 1496-008607, 1496-008615, 1496-008616, 1496-008617, 1496-008619, 1496-008621, 1496-008622, 1496-008623, 1496-009099, 1496-009101, 1496-009103, 1496-011913, 1496-012487, 1496-015399, 1496-016027, 1496-021455, 1496-023173, 1496-023211, 1496-023228, 1496-023489, 1496-033426, 1496-034558, 1496-034934, 1496-034941, 1496-035662, 1496-036272, 1496-036276, 1496-036456, 1496-036462, 1496-037647, 1496-044854, 1496-044856, 1496-044859, 1496-044861, 1496-044864, 1496-044865, 1496-044867, 1496-044868, 1496-044869, 1496-044870, 1496-044871, 1496-044874, 1496-044875, 1496-044877, 1496-044880, 1496-044881, 1496-044883, 1496-044887, 1496-044889, 1496-044893, 1496-044896, 1496-044898, 1496-044900, 1496-044901, 1496-044905, 1496-044906, 1496-044912, 1496-044962, 1496-044965, 1496-044966, 1496-044968, 1496-044970, 1496-044972, 1496-100499, 1496-103756, 1496-103760, 1496-103761, 1496-103762, 1496-103763, 1496-103767, 1496-103769, 1496-103771, 1496-103772, 1496-103773, 1496-103774, 1496-103776, 1496-107453, 1496-111584, 1496-111592, 1496-111602, 1496-111959, 1496-115116, 1496-115570, 1496-116848, 1496-120321, 1496-120346, 1496-120348, 1496-120351, 1496-120352, 1496-120928, 1496-121757, 1496-126546, 1496-126547, 1496-126548, 1496-126549, 1496-126550, 1496-126551, 1496-126552, 1496-126553, 1496-126554, 1496-126555, 1496-126556, 1496-126557, 1496-126558, 1496-126559, 1496-126560, 1496-126561, 1496-126562, 1496-126563, 1496-126564, 1496-126565, 1496-126566, 1496-126567, 1496-126568, 1496-128035, 1496-128202, 1496-128203, 1496-128204, 1496-128205, 1496-128206, 1496-128207, 1496-128208, 1496-128209, 1496-128210, 1496-128211, 1496-128212, 1496-128213, 1496-128214, 1496-128217, 1496-128218, 1496-128219, 1496-128220, 1496-128221, 1496-128222, 1496-128223, 1496-128224, 1496-128225, 1496-128226, 1496-128227, 1496-128228, 1496-128229, 1496-128230, 1496-128231, 1496-128232, 1496-128233, 1496-128234, 1496-128250, 1496-128251, 1496-128281, 1496-128282, 1496-128283, 1496-128284, 1496-128285, 1496-128287, 1496-128288, 1496-128289, 1496-128290, 1496-128291, 1496-128292, 1496-128293, 1496-128294, 1496-128295, 1496-128296, 1496-128297, 1496-128298, 1496-128299, 1496-128300, 1496-128301, 1496-128302, 1496-128303, 1496-129382, 1496-129383, 1496-129384, 1496-129385, 1496-129386, 1496-129387, 1496-129485, 1496-129486, 1496-129487, 1496-129488, 1496-129489, 1496-129490, 1496-129491, 1496-129492, 1496-129493, 1496-129494, 1496-129573, 1496-129637, 1496-129639, 1496-129640, 1496-129641, 1496-129642, 1496-129643, 1496-129644, 1496-129645, 1496-129646, 1496-129647, 1496-129648, 1496-129649, 1496-129650, 1496-129651, 1496-129652, 1496-129653, 1496-129654, 1496-129655, 1496-129656, 1496-129657, 1496-129658, 1496-129659, 1496-129660, 1496-129661, 1496-129662, 1496-129663, 1496-129664, 1496-129665, 1496-129666, 1496-129667, 1496-129668, 1496-129669, 1496-129670, 1496-129671, 1496-129672, 1496-129673, 1496-129674, 1496-129675, 1496-129676, 1496-129677, 1496-129678, 1496-129679, 1496-129680, 1496-129681, 1496-129682, 1496-129683, 1496-129684, 1496-129685, 1496-129686, 1496-129687, 1496-129688, 1496-129689, 1496-129690, 1496-129691, 1496-129692, 1496-129693, 1496-129694, 1496-129695, 1496-129696, 1496-129697, 1496-129698, 1496-129699, 1496-129700, 1496-129701, 1496-129702, 1496-129703, 1496-129704, 1496-129705, 1496-129706, 1496-129707, 1496-129708, 1496-129709, 1496-129710, 1496-129711, 1496-129712, 1496-129713, 1496-129714, 1496-129715, 1496-129865, 1496-129866, 1496-129867, 1496-129868, 1496-129869, 1496-129870, 1496-129871, 1496-129872, 1496-129873, 1496-129874, 1496-129875, 1496-129876, 1496-129877, 1496-129878, 1496-129879, 1496-129880, 1496-129881, 1496-129882, 1496-129887, 1496-129888, 1496-129889, 1496-129890, 1496-129891, 1496-129892, 1496-129893, 1496-129894, 1496-129895, 1496-129896, 1496-129897, 1496-129898, 1496-129899, 1496-129900, 1496-129901, 1496-129902, 1496-129903, 1496-129904, 1496-129905, 1496-129906, 1496-129907, 1496-129908, 1496-129909, 1496-129910, 1496-129911, 1496-129912, 1496-129913, 1496-129914, 1496-129915, 1496-129916, 1496-129917, 1496-129918, 1496-129919, 1496-129920, 1496-129921, 1496-129922, 1496-129923, 1496-129924, 1496-129925, 1496-129926, 1496-129927, 1496-129928, 1496-129929, 1496-129930, 1496-129931, 1496-129932, 1496-129933, 1496-129934, 1496-129935, 1496-129936, 1496-129937, 1496-129938, 1496-129939, 1496-129940, 1496-129941, 1496-129942, 1496-129943, 1496-129944, 1496-129945, 1496-129946, 1496-129947, 1496-129948, 1496-129949, 1496-129950, 1496-129951, 1496-129952, 1496-129953, 1496-129954, 1496-129955, 1496-129956, 1496-129957, 1496-129978, 1496-129979, 1496-129980, 1496-129981, 1496-129982, 1496-129983, 1496-129984, 1496-129985, 1496-129986, 1496-129987, 1496-129988, 1496-129989, 1496-129990, 1496-129991, 1496-129992, 1496-129993, 1496-129994, 1496-129995, 1496-129996, 1496-129997, 1496-129998, 1496-129999, 1496-130000, 1496-130001, 1496-130002, 1496-130003, 1496-130004, 1496-130005, 1496-130006, 1496-130007, 1496-130008, 1496-130009, 1496-130010, 1496-130011, 1496-130012, 1496-130013, 1496-130014, 1496-130015, 1496-130016, 1496-130017, 1496-130018, 1496-130019, 1496-130020, 1496-130021, 1496-130022, 1496-130023, 1496-130024, 1496-130025, 1496-130026, 1496-130027, 1496-130028, 1496-130029, 1496-130030, 1496-130031, 1496-130032, 1496-130033, 1496-130034, 1496-130035, 1496-130036, 1496-130037, 1496-130038, 1496-130039, 1496-130040, 1496-130041, 1496-130058, 1496-130059, 1496-130062, 1496-130063, 1496-130064, 1496-130065, 1496-130066, 1496-130067, 1496-130068, 1496-130069, 1496-130070, 1496-130071, 1496-130072, 1496-130073, 1496-130084, 1496-130085, 1496-130086, 1496-130087, 1496-130088, 1496-130089, 1496-130090, 1496-130091, 1496-130092, 1496-130093, 1496-130094, 1496-130095, 1496-130096, 1496-130097, 1496-130098, 1496-130099, 1496-130100, 1496-130101, 1496-130102, 1496-130103, 1496-130112, 1496-130113, 1496-130114, 1496-130115, 1496-130116, 1496-130117, 1496-130118, 1496-130119, 1496-130120, 1496-130121, 1496-130122, 1496-130123, 1496-130124, 1496-130125, 1496-130126, 1496-130127, 1496-130128, 1496-130129, 1496-130130, 1496-130131, 1496-130132, 1496-130133, 1496-130134, 1496-130135, 1496-130136, 1496-130137, 1496-130138, 1496-130139, 1496-130140, 1496-130141, 1496-130142, 1496-130143, 1496-130144, 1496-130145, 1496-130146, 1496-130147, 1496-130148, 1496-130149, 1496-130150, 1496-130151, 1496-130152, 1496-130153, 1496-130154, 1496-130155, 1496-130156, 1496-130157, 1496-130158, 1496-130159, 1496-130160, 1496-130161, 1496-130162, 1496-130163, 1496-130164, 1496-130165, 1496-130166, 1496-130167, 1496-130168, 1496-130169, 1496-130170, 1496-130171, 1496-130172, 1496-130173, 1496-130174, 1496-130175, 1496-130176, 1496-130177, 1496-130178, 1496-130179, 1496-130180, 1496-130181, 1496-130182, 1496-130183, 1496-130184, 1496-130185, 1496-130186, 1496-130187, 1496-130188, 1496-130189, 1496-130190, 1496-130191, 1496-130192, 1496-130193, 1496-130194, 1496-130195, 1496-130196, 1496-130197, 1496-130198, 1496-130199, 1496-130200, 1496-130201, 1496-130202, 1496-130203, 1496-130204, 1496-130205, 1496-130206, 1496-130207, 1496-130208, 1496-130209, 1496-130210, 1496-130211, 1496-130212, 1496-130213, 1496-130214, 1496-130215, 1496-130216, 1496-130217, 1496-130218, 1496-130219, 1496-130220, 1496-130221, 1496-130222, 1496-130223, 1496-130224, 1496-130225, 1496-130226, 1496-130227, 1496-130228, 1496-130229, 1496-130230, 1496-130231, 1496-130232, 1496-130233, 1496-130234, 1496-130235, 1496-130236, 1496-130237, 1496-130238, 1496-130239, 1496-130240, 1496-130241, 1496-130242, 1496-130243, 1496-130244, 1496-130245, 1496-130246, 1496-130247, 1496-130248, 1496-130249, 1496-130250, 1496-130251, 1496-130252, 1496-130253, 1496-130254, 1496-130255, 1496-130256, 1496-130257, 1496-130258, 1496-130259, 1496-130260, 1496-130261, 1496-130262, 1496-130263, 1496-130264, 1496-130265, 1496-130266, 1496-130267, 1496-130268, 1496-130269, 1496-130270, 1496-130271, 1496-130272, 1496-130273, 1496-130274, 1496-130275, 1496-130276, 1496-130277, 1496-130278, 1496-130279, 1496-130280, 1496-130281, 1496-130282, 1496-130283, 1496-130284, 1496-130286, 1496-130287, 1496-130288, 1496-130289, 1496-130290, 1496-130291, 1496-130292, 1496-130293, 1496-130294, 1496-130295, 1496-130303, 1496-130304, 1496-130305, 1496-130404, 1496-130405, 1496-130406, 1496-130407, 1496-130408, 1496-130409, 1496-130410, 1496-130411, 1496-130412, 1496-130413, 1496-130414, 1496-130415, 1496-130416, 1496-130417, 1496-130418, 1496-130419, 1496-130420, 1496-130421, 1496-130422, 1496-130423, 1496-130425, 1496-130426, 1496-130427, 1496-130428, 1496-130429, 1496-130430, 1496-130431, 1496-130432, 1496-130433, 1496-130434, 1496-130751, 1496-130752, 1496-130753, 1496-130754, 1496-130755, 1496-130756, 1496-130757, 1496-130758, 1496-130759, 1496-130760, 1496-130761, 1496-130767, 1496-130772, 1496-130863, 1496-130864, 1496-130865, 1496-130866, 1496-130867, 1496-130868, 1496-130869, 1496-130870, 1496-130871, 1496-130872, 1496-130875, 1496-130876, 1496-130877, 1496-130878, 1496-130879, 1496-131040, 1496-131041, 1496-131042, 1496-131043, 1496-131044, 1496-131045, 1496-131046, 1496-131047, 1496-131048, 1496-131049, 1496-131050, 1496-131071, 1496-131072, 1496-131073, 1496-131074, 1496-131075, 1496-131076, 1496-131077, 1496-131078, 1496-131079, 1496-131080, 1496-131081, 1496-131083, 1496-131126, 1496-131127, 1496-131128, 1496-131129, 1496-131130, 1496-131131, 1496-131132, 1496-131133, 1496-131134, 1496-131163, 1496-131164, 1496-131165, 1496-131166, 1496-131167, 1496-131168, 1496-131169, 1496-131170, 1496-131171, 1496-131172, 1496-131173, 1496-131193, 1496-131194, 1496-131195, 1496-131196, 1496-131197, 1496-131198, 1496-131199, 1496-131200, 1496-131201, 1496-131202, 1496-131601, 1496-131617, 1496-131618, 1496-131619, 1496-131620, 1496-131621, 1496-131622, 1496-131623, 1496-131624, 1496-131625, 1496-131627, 1496-131628, 1496-131629, 1496-131630, 1496-131631, 1496-131632, 1496-131633, 1496-131634, 1496-131635, 1496-131636, 1496-131637, 1496-131638, 1496-131639, 1496-131640, 1496-131641, 1496-131642, 1496-131643, 1496-131644, 1496-131690, 1496-131691, 1496-131692, 1496-131693, 1496-131694, 1496-131695, 1496-131712, 1496-131713, 1496-131714, 1496-131715, 1496-131716, 1496-131717, 1496-131718, 1496-131719, 1496-131720, 1496-131721, 1496-131722, 1496-131723, 1496-131724, 1496-131725, 1496-131726, 1496-131727, 1496-131728, 1496-131729, 1496-131730, 1496-131731, 1496-131732, 1496-131733, 1496-131734, 1496-131735, 1496-131736, 1496-131737, 1496-131738, 1496-131739, 1496-131740, 1496-131741, 1496-131742, 1496-131743, 1496-131744, 1496-131745, 1496-131746, 1496-131747, 1496-131780, 1496-131781, 1496-131782, 1496-131783, 1496-131944, 1496-131945, 1496-131946, 1496-131947, 1496-131948, 1496-131949, 1496-131950, 1496-131951, 1496-131952, 1496-131953, 1496-131954, 1496-131955, 1496-131956, 1496-131957, 1496-131958, 1496-131959, 1496-131960, 1496-131961, 1496-131962, 1496-131963, 1496-131964, 1496-131965, 1496-131966, 1496-131967, 1496-131968, 1496-131969, 1496-132217, 1496-132218, 1496-132219, 1496-132220, 1496-132221, 1496-132222, 1496-132223, 1496-132224, 1496-132225, 1496-132227 , 1496-132228 , 1496-132229, 1496-132230, 1496-132236, 1496-132237, 1496-132238, 1496-132239, 1496-132240, 1496-132241, 1496-132242, 1496-132243, 1496-132244, 1496-132245, 1496-132246, 1496-132247, 1496-132248, 1496-132249, 1496-132250, 1496-132252, 1496-132376, 1496-132377, 1496-132378, 1496-132379, 1496-132380, 1496-132381, 1496-132382, 1496-132924, 1496-132925, 1496-132926, 1496-132927, 1496-132928, 1496-132929, 1496-132930, 1496-132931, 1496-132932, 1496-132933, 1496-132934, 1496-132935, 1496-132936, 1496-132937, 1496-132938, 1496-132939, 1496-132940, 1496-132941, 1496-132942, 1496-132943, 1496-133145, 1496-133146, 1496-133147, 1496-133148, 1496-133149, 1496-133150, 1496-133151, 1496-133152, 1496-133153, 1496-133161, 1496-133178, 1496-133179, 1496-133180, 1496-133181, 1496-133182, 1496-133183, 1496-133184, 1496-133185, 1496-133186, 1496-133187, 1496-133188, 1496-133189, 1496-133190, 1496-133191, 1496-133192, 1496-133193, 1496-133194, 1496-133195, 1496-133196, 1496-133197, 1496-133198, 1496-133199, 1496-133200, 1496-133201, 1496-133202, 1496-133203, 1496-133204, 1496-133205, 1496-133206, 1496-133207, 1496-133350, 1496-133351, 1496-133352, 1496-133353, 1496-133354, 1496-133355, 1496-133356, 1496-133357, 1496-133358, 1496-133359, 1496-133360, 1496-133361, 1496-133362, 1496-133363, 1496-133364, 1496-133365, 1496-133366, 1496-133367, 1496-133368, 1496-133369, 1496-133370, 1496-133371, 1496-133372, 1496-133373, 1496-133374, 1496-133375, 1496-133376, 1496-133377, 1496-133378, 1496-133379, 1496-133380, 1496-133381, 1496-133382, 1496-133383, 1496-133384, 1496-133385, 1496-133386, 1496-133387, 1496-133388, 1496-133389, 1496-133390, 1496-133391, 1496-133392, 1496-133393, 1496-133394, 1496-133395, 1496-133396, 1496-133397, 1496-133398, 1496-133399, 1496-133400, 1496-133401, 1496-133402, 1496-133403, 1496-133404, 1496-133405, 1496-133406, 1496-133407, 1496-133408, 1496-133409, 1496-133410, 1496-133411, 1496-133412, 1496-133413, 1496-133414, 1496-133415, 1496-133416, 1496-133417, 1496-133418, 1496-133419, 1496-133420, 1496-133421, 1496-133422, 1496-133423, 1496-133424, 1496-133425, 1496-133426, 1496-133427, 1496-133428, 1496-133429, 1496-133430, 1496-133431, 1496-133432, 1496-133433, 1496-133434, 1496-133435, 1496-133436, 1496-133437, 1496-133438, 1496-133439, 1496-133440, 1496-133441, 1496-133442, 1496-133443, 1496-133444, 1496-133445, 1496-133446, 1496-133447, 1496-133448, 1496-133449, 1496-133450, 1496-133451, 1496-133452, 1496-133453, 1496-133454, 1496-133455, 1496-133456, 1496-133457, 1496-133458, 1496-133459, 1496-133460, 1496-133492, 1496-133600, 1496-133601, 1496-133602, 1496-133603, 1496-133604, 1496-133605, 1496-133606, 1496-133607, 1496-133608, 1496-133609, 1496-133611, 1496-133612, 1496-133613, 1496-133614, 1496-133615, 1496-133616, 1496-133617, 1496-133618, 1496-133619, 1496-133620, 1496-133621, 1496-133622, 1496-133623, 1496-133628, 1496-133629, 1496-133630, 1496-133631, 1496-133632, 1496-133633, 1496-133634, 1496-133635, 1496-133636, 1496-133637, 1496-133638, 1496-133639, 1496-133640, 1496-133641, 1496-133642, 1496-133643, 1496-133644, 1496-133645, 1496-133646, 1496-133647, 1496-133648, 1496-133649, 1496-133650, 1496-133651, 1496-133652, 1496-133653, 1496-133654, 1496-133655, 1496-133656, 1496-133657, 1496-133658, 1496-133659, 1496-133660, 1496-133661, 1496-133662, 1496-133663, 1496-133664, 1496-133665, 1496-133666, 1496-133667, 1496-133668, 1496-133669, 1496-133670, 1496-133675, 1496-133676, 1496-133677, 1496-133678, 1496-133887, 1496-133888, 1496-133889, 1496-133890, 1496-133891, 1496-133892, 1496-133893, 1496-133894, 1496-133895, 1496-133896, 1496-133980, 1496-133981, 1496-133982, 1496-133983, 1496-133984, 1496-133985, 1496-133986, 1496-133987, 1496-133988, 1496-133989, 1496-133990, 1496-133991, 1496-133992, 1496-133993, 1496-133994, 1496-133995, 1496-133996, 1496-133997, 1496-133998, 1496-133999, 1496-134000, 1496-134001, 1496-134002, 1496-134003, 1496-134004, 1496-134005, 1496-134006, 1496-134007, 1496-134008, 1496-134009, 1496-134010, 1496-134011, 1496-134012, 1496-134013, 1496-134014, 1496-134015, 1496-134016, 1496-134017, 1496-134018, 1496-134019, 1496-134020, 1496-134021, 1496-134022, 1496-134023, 1496-134024, 1496-134025, 1496-134026, 1496-134027, 1496-134028, 1496-134029, 1496-134030, 1496-134031, 1496-134032, 1496-134048, 1496-134049, 1496-134050, 1496-134051, 1496-134052, 1496-134053, 1496-134054, 1496-134055, 1496-134056, 1496-134057, 1496-134058, 1496-134059, 1496-134060, 1496-134061, 1496-134126, 1496-134127, 1496-134128, 1496-134179, 1496-134180, 1496-134181, 1496-134182, 1496-134183, 1496-134184, 1496-134185, 1496-134186, 1496-134187, 1496-134188, 1496-134189, 1496-134190, 1496-134191, 1496-134192, 1496-134193, 1496-134194, 1496-134430, 1496-134431, 1496-134432, 1496-134433, 1496-134434, 1496-134435, 1496-134436, 1496-134437, 1496-134438, 1496-134439, 1496-134440, 1496-134441, 1496-134442, 1496-134443, 1496-134482, 1496-134484, 1496-134485, 1496-134486, 1496-134487, 1496-134488, 1496-134489, 1496-134490, 1496-134491, 1496-134492, 1496-134493, 1496-134494, 1496-134495, 1496-134496, 1496-134497, 1496-134498, 1496-134499, 1496-134500, 1496-134501, 1496-134502, 1496-134503, 1496-134504, 1496-134505, 1496-134506, 1496-134507, 1496-134508, 1496-134509, 1496-134510, 1496-134511, 1496-134512, 1496-134513, 1496-134514, 1496-134515, 1496-134516, 1496-134517, 1496-134518, 1496-134519, 1496-134726, 1496-134727, 1496-134728, 1496-134729, 1496-134730, 1496-134731, 1496-134732, 1496-134733, 1496-134734, 1496-134735, 1496-134736, 1496-134737, 1496-134738, 1496-134739, 1496-134740, 1496-134741, 1496-134742, 1496-134743, and 1496-134744. **********SERIAL NUMBERS OF MODULES OUTSIDE THE US****************** 1496-033286, 1496-126260, 1496-126277, 1496-126278, 1496-126279, 1496-126280, 1496-126281, 1496-128027, 1496-128045, 1496-128046, 1496-128051, 1496-128153, 1496-128154, 1496-128155, 1496-128156, 1496-128157, 1496-128158, 1496-128159, 1496-128160, 1496-128161, 1496-128162, 1496-128163, 1496-128164, 1496-128165, 1496-128166, 1496-128167, 1496-128168, 1496-128169, 1496-128170, 1496-128171, 1496-128172, 1496-128173, 1496-128174, 1496-128175, 1496-128176, 1496-128177, 1496-128178, 1496-128179, 1496-128180, 1496-128181, 1496-128182, 1496-128183, 1496-128184, 1496-128197, 1496-128198, 1496-128199, 1496-128200, 1496-128201, 1496-128215, 1496-128216, 1496-129066, 1496-129067, 1496-129068, 1496-129069, 1496-129070, 1496-129379, 1496-129380, 1496-129381, 1496-129883, 1496-129884, 1496-129885, 1496-129886, 1496-130042, 1496-130043, 1496-130044, 1496-130045, 1496-130046, 1496-130047, 1496-130048, 1496-130049, 1496-130050, 1496-130051, 1496-130052, 1496-130053, 1496-130054, 1496-130055, 1496-130056, 1496-130057, 1496-130060, 1496-130061, 1496-130074, 1496-130075, 1496-130076, 1496-130077, 1496-130078, 1496-130079, 1496-130080, 1496-130081, 1496-130285, 1496-130296, 1496-130297, 1496-130298, 1496-130299, 1496-130300, 1496-130301, 1496-130302, 1496-130424, 1496-130435, 1496-130436, 1496-130437, 1496-130438, 1496-130439, 1496-130440, 1496-130441, 1496-130442, 1496-130443, 1496-130444, 1496-130445, 1496-130446, 1496-130762, 1496-130763, 1496-130764, 1496-130765, 1496-130766, 1496-130768, 1496-130769, 1496-130770, 1496-130771, 1496-130773, 1496-130774, 1496-130775, 1496-130776, 1496-130777, 1496-130778, 1496-130779, 1496-130780, 1496-130788, 1496-130789, 1496-130790, 1496-130791, 1496-130792, 1496-130793, 1496-130794, 1496-130795, 1496-130796, 1496-130797, 1496-130798, 1496-130799, 1496-130800, 1496-130801, 1496-130802, 1496-130803, 1496-130804, 1496-130805, 1496-130806, 1496-130807, 1496-130808, 1496-130809, 1496-130810, 1496-130811, 1496-130812, 1496-130813, 1496-130814, 1496-130815, 1496-130816, 1496-130817, 1496-130818, 1496-130819, 1496-130820, 1496-130821, 1496-130822, 1496-130823, 1496-130824, 1496-130825, 1496-130826, 1496-130827, 1496-130828, 1496-130829, 1496-130830, 1496-130831, 1496-130832, 1496-130833, 1496-130834, 1496-130835, 1496-130836, 1496-130837, 1496-130838, 1496-130839, 1496-130840, 1496-130841, 1496-130842, 1496-130843, 1496-130844, 1496-130845, 1496-130846, 1496-130847, 1496-130848, 1496-130849, 1496-130850, 1496-130851, 1496-130852, 1496-130853, 1496-130854, 1496-130855, 1496-130856, 1496-130857, 1496-130858, 1496-130859, 1496-130860, 1496-130861, 1496-130862, 1496-130873, 1496-130874, 1496-130914, 1496-130915, 1496-130916, 1496-130917, 1496-130918, 1496-130919, 1496-130920, 1496-131030, 1496-131031, 1496-131032, 1496-131033, 1496-131034, 1496-131035, 1496-131036, 1496-131037, 1496-131038, 1496-131039, 1496-131082, 1496-131084, 1496-131085, 1496-131086, 1496-131087, 1496-131088, 1496-131089, 1496-131090, 1496-131125, 1496-131154, 1496-131155, 1496-131156, 1496-131157, 1496-131158, 1496-131159, 1496-131160, 1496-131161, 1496-131162, 1496-131433, 1496-131434, 1496-131435, 1496-131436, 1496-131437, 1496-131438, 1496-131439, 1496-131440, 1496-131441, 1496-131442, 1496-131443, 1496-131444, 1496-131445, 1496-131446, 1496-131447, 1496-131448, 1496-131449, 1496-131450, 1496-131451, 1496-131452, 1496-131592, 1496-131593, 1496-131594, 1496-131595, 1496-131596, 1496-131597, 1496-131598, 1496-131600, 1496-131626, 1496-131645, 1496-131646, 1496-131647, 1496-131648, 1496-131649, 1496-131650, 1496-131651, 1496-131652, 1496-131653, 1496-131654, 1496-131655, 1496-131656, 1496-131657, 1496-131658, 1496-131659, 1496-131660, 1496-131661, 1496-131662, 1496-131663, 1496-131664, 1496-131665, 1496-131666, 1496-131667, 1496-131668, 1496-131669, 1496-131670, 1496-131671, 1496-131672, 1496-131673, 1496-131674, 1496-131675, 1496-131676, 1496-131677, 1496-131678, 1496-131679, 1496-131680, 1496-131681, 1496-131682, 1496-131683, 1496-131684, 1496-131685, 1496-131686, 1496-131687, 1496-131688, 1496-131689, 1496-131696, 1496-131697, 1496-131707, 1496-131708, 1496-131709, 1496-131710, 1496-131711, 1496-132253, 1496-132265, 1496-132266, 1496-132267, 1496-132268, 1496-132269, 1496-132270, 1496-132271, 1496-132272, 1496-132273, 1496-132274, 1496-132275, 1496-132276, 1496-132277, 1496-132278, 1496-132279, 1496-132280, 1496-132281, 1496-132282, 1496-132283, 1496-132284, 1496-132358, 1496-132359, 1496-132360, 1496-132361, 1496-132362, 1496-132363, 1496-132364, 1496-132365, 1496-132366, 1496-132367, 1496-132368, 1496-132369, 1496-132370, 1496-132371, 1496-132372, 1496-132373, 1496-132374, 1496-132375, 1496-132923, 1496-132944, 1496-133133, 1496-133134, 1496-133135, 1496-133136, 1496-133137, 1496-133138, 1496-133139, 1496-133140, 1496-133141, 1496-133142, 1496-133143, 1496-133144, 1496-133154, 1496-133155, 1496-133156, 1496-133157, 1496-133158, 1496-133159, 1496-133160, 1496-133162, 1496-133163, 1496-133164, 1496-133165, 1496-133166, 1496-133167, 1496-133168, 1496-133169, 1496-133170, 1496-133171, 1496-133172, 1496-133173, 1496-133174, 1496-133175, 1496-133176, 1496-133177, 1496-133461, 1496-133483, 1496-133484, 1496-133485, 1496-133486, 1496-133487, 1496-133488, 1496-133489, 1496-133490, 1496-133491, 1496-133493, 1496-133494, 1496-133495, 1496-133496, 1496-133497, 1496-133498, 1496-133499, 1496-133595, 1496-133596, 1496-133597, 1496-133598, 1496-133599, 1496-133610, 1496-133624, 1496-133625, 1496-133626, 1496-133627, 1496-133671, 1496-133672, 1496-133673, 1496-133674, 1496-133947, 1496-133948, 1496-133949, 1496-133950, 1496-133951, 1496-133952, 1496-133953, 1496-133954, 1496-133955, 1496-133956, 1496-133957, 1496-133958, 1496-133959, 1496-133960, 1496-133961, 1496-133962, 1496-133963, 1496-133964, 1496-133965, 1496-133966, 1496-133967, 1496-133968, 1496-133969, 1496-133970, 1496-133971, 1496-133972, 1496-133973, 1496-133974, 1496-133975, 1496-133976, 1496-133977, 1496-133978, 1496-133979, 1496-134033, 1496-134034, 1496-134035, 1496-134036, 1496-134037, 1496-134038, 1496-134039, 1496-134040, 1496-134041, 1496-134042, 1496-134043, 1496-134044, 1496-134045, 1496-134046, 1496-134047, 1496-134195, 1496-134196, 1496-134197, 1496-134198, 1496-134199, 1496-134200, 1496-134201, 1496-134202, 1496-134203, 1496-134204, 1496-134205, 1496-134351, 1496-134352, 1496-134353, 1496-134354, 1496-134355, 1496-134356, 1496-134357, 1496-134358, 1496-134359, 1496-134360, 1496-134361, 1496-134362, 1496-134363, 1496-134364, 1496-134365, 1496-134366, 1496-134367, 1496-134368, 1496-134369, 1496-134370, 1496-134371, 1496-134372, 1496-134373, 1496-134374, 1496-134375, 1496-134376, 1496-134377, 1496-134378, 1496-134379, 1496-134380, 1496-134381, 1496-134382, 1496-134383, 1496-134384, 1496-134385, 1496-134386, 1496-134387, 1496-134388, 1496-134389, 1496-134390, 1496-134391, 1496-134392, 1496-134393, 1496-134394, 1496-134395, 1496-134396, 1496-134397, 1496-134398, 1496-134399, 1496-134400, 1496-134401, 1496-134402, 1496-134403, 1496-134404, 1496-134405, 1496-134406, 1496-134407, 1496-134408, 1496-134409, 1496-134410, 1496-134411, 1496-134412, 1496-134413, 1496-134414, 1496-134415, 1496-134416, 1496-134417, 1496-134418, 1496-134419, 1496-134420, 1496-134421, 1496-134422, 1496-134423, 1496-134424, 1496-134425, 1496-134426, 1496-134427, 1496-134428, 1496-134429, 1496-134458, 1496-134459, 1496-134460, 1496-134461, 1496-134462, 1496-134463, 1496-134464, 1496-134465, 1496-134466, 1496-134467, 1496-134468, 1496-134469, 1496-134470, 1496-134471, 1496-134472, 1496-134473, 1496-134474, 1496-134475, 1496-134476, 1496-134477, 1496-134478, 1496-134479, 1496-134480, 1496-134481, 1496-134483, 1496-134638, 1496-134639, 1496-134640, 1496-134641, 1496-134642, 1496-134643, 1496-134644, 1496-134645, 1496-134646, 1496-134647, 1496-134648, 1496-134649, 1496-134650, 1496-134651, and 1496-134751.
Other devices under the same FDA submission
2 devices cleared under K103142 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Ultraview SL Ultraview SL®Spacelabs Healthcare (Washington), Inc · SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE2 recalls2 reports
- Ultraview SL UV SL Command ModuleSpacelabs Healthcare (Washington), Inc · DETECTOR AND ALARM, ARRHYTHMIA1 recall3 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.