Skip to content

Tapered HD Upgrade Set, REF TSKHDUS, Non-Sterile, Rx Only recallZ-1605-2015 · BioHorizons Implant Systems Inc · initiated Apr 16, 2015

Class II Terminated All recalls posted in May 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from BioHorizons Implant Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BioHorizons Tapered HD Upgrade set was packaged with an incorrect drill component manufactured of stainless steel in place of the correct component manufactured from titanium.

FDA root cause category: Process control

What the company is doing

BioHorizons sent an Urgent Medical Device Field Correction letter on 04/16/2015 to affected customers. Non-responding customers will be notified after 30 days. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Care Department at 888-246-8338.

Recall details

Recall number
Z-1605-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 16, 2015
Posted by the FDA
May 12, 2015
Recalling firm
BioHorizons Implant Systems Inc
Quantity
48 units
Product code
DZE
FDA submissions
K071638
Distribution
US Nationwide in the states of: AL, FL, MS, VA, LA, CA, NC, TX, TN, OH, OK, VT, WA, MD, NY, SD, and SC.

Product description

Tapered HD Upgrade Set, REF TSKHDUS, Non-Sterile, Rx Only, Product Usage Usage: Surgical instruments used in preparation for dental implant placement

Code information (models, lots, serial numbers)

Lot Number 1501968

Other devices under the same FDA submission

121 devices cleared under K071638 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page