Dade Actin Activated Cephaloplastin Reagent recallZ-1703-2015 · Siemens Healthcare Diagnostics, Inc. · initiated Mar 25, 2015

Class II Terminated All recalls posted in June 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Recovery for assayed controls are out of assigned ranges. A drift >3 seconds in the normal APTT range and/or >15% in the pathological range has been observed. Patient samples with recoveries close to the medical decision points could exhibit a deviation up to 4 seconds for the normal range and up to 33% for the pathological range.

FDA root cause category: Under Investigation by firm

What the company is doing

On March 24, 2015 an Urgent Medical Device Recall letter was mailed to all affected customers to notify them about the issue, associated risks, and actions to be taken.

Recall details

Recall number
Z-1703-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 25, 2015
Posted by the FDA
Jun 2, 2015
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
27501
Product code
GFO
FDA submissions
K760318
Distribution
Worldwide Distribution. US naionwide including Puerto Rico, Barbados and Mexico.

Product description

Dade Actin Activated Cephaloplastin Reagent; Liquid rabbit brain cephalin with plasma activator for use in the determination of the activated partial thromboplastin time (APTT) and other coagulation procedures requiring an activated partial thromboplastin reagent.

Code information (models, lots, serial numbers)

Catalog Number B4218-1, Material Number 10445709, lot numbers 557128, 557129, 557130, 557131, 557132, 557133, 557134, 557135, 557140, 557141 , expiration dates:10/31/2015, 11/21/2015, 2/16/2016

Other devices under the same FDA submission

1 device cleared under K760318 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.