Skip to content

SOMATOM Perspective recallZ-1646-2015 · Siemens Medical Solutions USA, Inc · initiated Apr 20, 2015

Class II Terminated All recalls posted in May 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software bugs in VC20B SP0a or SP1 software versions may cause issues that could make it necessary to rescan patients. Syngo Main UI may crash if the Patient Browser is scrolled with arrow keys of the keyboard. Tomo images may freeze on the screen when zoomed in or out under the mode of CAREVision. Sporadic displaying error in WorkStream4D application. DB may lock under heavy and multitasks.

FDA root cause category: Software design

What the company is doing

Siemens sent an Customer Safety Advisory Notice dated April 20, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A remote software update is scheduled for release that will resolve the issue. For further questions, please call (610) 219-6300.

Recall details

Recall number
Z-1646-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 20, 2015
Posted by the FDA
May 22, 2015
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
75
Product code
JAK
FDA submissions
K133590
Distribution
Nationwide Distribution including Puerto Rico.

Product description

SOMATOM Perspective; the intended use of computed tomography is to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Code information (models, lots, serial numbers)

SOMATOM Perspective - 10495568  with serial numbers: 59019 59195 59060 59043 59025 59029 59189 59204 76954 59010 59164 59185 59048 59169 77107 77110 59011 59030 59171 59168 59017 59167 59162 59190 59047 59061 77121 59028 59165 59020 59181 59037 59035 59005 59183 59016 59027 59184 76950 59203 59015 77182

Other devices under the same FDA submission

3 devices cleared under K133590 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 71101).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page