Integra Camino Intracranial Pressure Monitoring Catheter… recallZ-1815-2015 · Integra LifeSciences Corp. · initiated May 27, 2015

Class II Terminated All recalls posted in June 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Integra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Catheters because they may have been incorrectly assembled. Specifically, the o-ring in the bolt adapter may be in the wrong location and as a result, there is a potential CSF/blood could leak from the bolt adapter.

FDA root cause category: Under Investigation by firm

What the company is doing

Integra sent an Urgent Medical Device Recall letter dated May 27, 2015, to all customers who received the Camino Intracranial Pressure Monitoring Catheters because they may have been incorrectly assembled. Specifically, the o-ring in the bolt adapter may be in the wrong location and as a result, there is a potential CSF/blood could leak from the bolt adapter. The letter informs the customers of the problems identified and the actions to be taken. Customers were instructed to complete the customer response form and return it by email or fax to: [email protected] or FAX to 1-609-275-9445. Customers with any questions regarding the instructions are instructed to contact Customer Service at 1-888-601-0203. For questions regarding this recall call 609-936-6822.

Recall details

Recall number
Z-1815-2015
Recall class
Class II
Status
Terminated
Initiated by the company
May 27, 2015
Posted by the FDA
Jun 18, 2015
Recalling firm
Integra LifeSciences Corp.
Quantity
1218 units
Product code
GWM
FDA submissions
K102875
Distribution
Worldwide Distribution - US (Nationwide) and Internationally to ARGENTINA, BRAZIL, BULGARIA, CANADA, COLUMBIA, COSTA RICA, FRANCE, GERMANY, ITALY, MEXICO, GERMANY, ITALY, MEXICO, PANAMA, SOUTH AFRICA, SPAIN, SWITZERLAND, UNITED ARAB EMIRATES, AND UNITED KINGDOM

Product description

Integra Camino Intracranial Pressure Monitoring Catheter with Licox IMC Bolt Fitting, Catalog No. 110-4L The Camino Intracranial Pressure Monitoring Catheter is a sterile transducer-tipped pressure monitoring catheter with accessory items to be used as a diagnostic tool for rapidly determining and continuously monitoring intracranial pressure. The Camino catheter has a miniature transducer at the distal tip. This design eliminates the need for a fluid-filled system to carry pressure waves to an external transducer and thus, the problems inherent in such fluid-filled systems are eliminated. The Camino catheter is packaged with a zero adjustment tool which allows for the adjustment of the transducer during catheter preparation, and the LICOX Bolt Adapter Fitting. The Camino Intracranial Pressure Monitoring Catheter with Licox IMC Bolt fitting, Model 110-4L is indicated for use by a qualified Neurosurgeon when direct measurement of intracranial pressure in the parenchyma is clinically important. It is intended to be used only through a Licox Brain Oxygen Monitoring System Bolt.

Code information (models, lots, serial numbers)

305000291459, 305000299228, 305000300396, 305000301413, 305000312261, 305000315716, 30500X294240, 30500X304371, 30500X309674, 30500X318076, 3050RX285879, 3050RY297365, 305E00318970, 305E0X320986

Other devices under the same FDA submission

3 devices cleared under K102875 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.