Skip to content

Vilex, Inc., FUZE 10.0 mm x 150 mm, Product # FZ100-150T-25 recallZ-1721-2015 · Vilex In Tennessee Inc · initiated Apr 22, 2015

Class II Terminated All recalls posted in June 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Vilex In Tennessee Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incidence of breakage is higher than expected.

FDA root cause category: Device Design

What the company is doing

The initial telephone call (4/9/2015) informed the distributor that the product was on hold (not to be used or transferred). The second telephone call (4/29/2015) and letter informed the customer that the product was being removed from market and that the customer was to return the product to the corporate office. The letter was disseminated on 05/05/2015 to distributors via certified mail. The letter contains the instructions for them to follow, which as summarized are: inventory all products on hand, sign off on inventory sheet, send products back to the corporate office with the identified return number on the shipping container, inform the corporate office that the process has been completed and provide the corporate office with the list of the products returned. A second letter was disseminated to sales representatives (distributors), surgeons, and hospitals on 05/28/2015.

Recall details

Recall number
Z-1721-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 22, 2015
Posted by the FDA
Jun 8, 2015
Recalling firm
Vilex In Tennessee Inc
Quantity
13 units
Product code
HSB
FDA submissions
K102413
Distribution
Distributed in the states of MS, MD, PA, and TX.

Product description

Vilex, Inc., FUZE 10.0 mm x 150 mm, Product # FZ100-150T-25. Ankle arthrodesis, tibio-talo-calcaneal arthrodesis.

Code information (models, lots, serial numbers)

all codes

Other devices under the same FDA submission

1 device cleared under K102413 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 71344).