Skip to content

Fortex" Pedicle Screw System Continuous Radius Rod 5.5 -… recallZ-1823-2015 · X Spine Systems Inc · initiated Apr 7, 2015

Class II Terminated All recalls posted in June 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from X Spine Systems Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Continuous Radius Rod should measure 50mm in length but actually measures 45mm in length.

FDA root cause category: Other

What the company is doing

X-Spine sent an Urgent Medical Device Recall Letter dated April 13, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Actions to be taken: 1. The affected hospital and surgeon should be notified of this recall immediately. 2. Complete the attached Medical Device Recall Return Response Form and return it with any affected product in inventory to X-Spine Systems immediately. a. Reference RMA # 5410 b. Use enclosed shipping label 3. No action should be taken with product that has been implanted. Please feel free to contact me at 800.903.0640 ext. 137.

Recall details

Recall number
Z-1823-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Apr 7, 2015
Posted by the FDA
Jun 22, 2015
Recalling firm
X Spine Systems Inc
Quantity
75 units
Product code
MNH
FDA submissions
K120832
Distribution
Worldwide Distribution - US Distribution to the states of : FL, ID, IL, KS, LA, MI, MO, NJ, OH, PA and TX., and to the countries of Austria, Germany and Italy.

Product description

Fortex" Pedicle Screw System Continuous Radius Rod 5.5 - 50mm, X-Spine Systems, Inc., 452 Alexandersville Road Miamisburg, OH 45342 Phone: 937-847-8400

Code information (models, lots, serial numbers)

Model #: N60001068; Lot #: 2943-01; Serial #s: 46118, 46581, 47296, 47588, 47653, 48164, 48201, 48412, 49156, 49585, 49586, 49649, G22, G45, G47, G77, G83, G87X, G90X, R15, XP03, XP21, XP27, XP28, XP29 & XP32.

Other devices under the same FDA submission

5 devices cleared under K120832 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page