IMMULITE 2000/IMMULITE 2000 XPi Italian Cypress 3g Allergy… recallZ-1843-2015 · Siemens Healthcare Diagnostics, Inc. · initiated May 6, 2015
Class II Terminated All recalls posted in June 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
IMMULITE 2000/2000XPi Italian Cypress (T23L4), Lots 206 and 208, fail to meet specifications to support the shelf-life stability of 2 years (24 months) expiration date as indicated on the products label.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens Healthcare Diagnostics issued an Urgent Medical Device Recall Letter (IMC-15-11A.US, dated May 6, 2015) in the US via FedEx, and an Urgent Field Safety Notice (IMC-15-11A.OUS, May 2015) outside the US on May 6, 2015. Customers were instructed to immediately discontinue use of and discard IMMULITE 2000/IMMULITE 2000 XPi Italian Cypress allergen, lot 206. The shelf-life of IMMULITE 2000/IMMULITE 2000 XPi Italian Cypress allergen, lot 208, has been shortened to June 30, 2015. Discontinue use and discard lot 208 after June 30, 2015. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Recall letter within 30 days in the US and OUS and fax to Customer Care Center at 312-275-7795.
Recall details
- Recall number
- Z-1843-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- May 6, 2015
- Posted by the FDA
- Jun 25, 2015
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- Domestic: 17 kits, Foreign: 1549 kits
- Product code
- DHB
- FDA submissions
- K013134
- Distribution
- Worldwide Distribution-Distributed in the states of Florida, New Mexico and New York and the countries of Australia, Belgium, Chile, Costa Rica, Croatia, Czech Republic, France, Germany, Greece, Guatemala, Israel, Italy, Japan, Kazakhstan, Latvia, Netherlands, Portugal, Romania, Russian Federation, Slovakia, and Spain.
Product description
IMMULITE 2000/IMMULITE 2000 XPi Italian Cypress 3g Allergy, Catalog # T23L4 (40 test), Siemens Material Number (SMN) 10386079, IVD. For in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers for the quantitative measurement of allergen-specific IgE in human serum, as an aid in the clinical diagnosis of IgE-mediated allergic disorders.
Code information (models, lots, serial numbers)
Kit lots 206 (exp. 6-30-2015) and 208 (exp. 11-30-2015)
Other devices under the same FDA submission
3 devices cleared under K013134 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- IMMULITE®2000 Systems 3gAllergy™ White Ash 10386060Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL4 reports
- IMMULITE®2000 Systems 3gAllergy™ Rat Epithelium 10385658Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL1 report
- IMMULITE®2000 Systems 3gAllergy™ Timothy Grass 10385924Siemens Healthcare Diagnostics Products Ltd · SYSTEM, TEST, RADIOALLERGOSORBENT (RAST) IMMUNOLOGICAL1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page