Medtronic DLP Femoral Arterial Cannula, models 57414… recallZ-1914-2015 · Medtronic Perfusion Systems · initiated Jun 12, 2015
Class II Terminated All recalls posted in June 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic is recalling select lots of DLP Femoral Arterial Cannula product due to recent reports in which the user had difficulty or was unable to connect the cannula to the perfusion circuit tubing.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Medtronic sent an Urgent Medical Device Recall letter, dated June 2015 to customers beginning June 12, 2015. The letter identified affected product, described the issue, requested that unused inventory be quarantined and returned to Medtronic. Customers should contact Customer Service at 800-854-3570 and reference RADAR #241958 to initiate a return and credit of product. Additionally a Customer Confirmation Certificate should be completed and faxed to 651-367-0612.
Recall details
- Recall number
- Z-1914-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 12, 2015
- Posted by the FDA
- Jun 26, 2015
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 1304 devices
- Product code
- DWF
- FDA submissions
- K875353
- Distribution
- Worldwide Distribution - US Distribution including the states of : AK, AZ, CA, CO, IN, KS, LA, MN, MT, NM, NY, OH, OR, PA, TX, UT, and WI., and to the countries of : Belgium, Brazil, Canada, Colombia, Denmark, France, Germany, Greece, Hungary, India, Ireland, Italy, Korea, Morocco, Netherlands, Poland, Qatar, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Turkey, and United Kingdom.
Product description
Medtronic DLP Femoral Arterial Cannula, models 57414, 57417, and 57421, and DLP Femoral Cannula Replacement Kit, Models 96017 and 96021. Sterilized using Ethylene Oxide. The cannula and dilator components are polyurethane and have tapered proximal tips to aid in diameter transition from the dilator. The dilator components utilize tip indicator bands for minimum tip alignment of the component parts. The cannula body is marked for depth insertion. The components are designed to fit over each other to progressively dilate the vessel during insertion. The dilator allows a 0.1-cm (0.038-in) guidewire to pass through it to aid in insertion. A soft, tapered transition fitting terminates in a 0.95-cm (3/8-in) connection site or barbed connector.
Code information (models, lots, serial numbers)
Model 57414 / Lot Numbers: 2014124994, 2015016150, 2015016729, 2015020381, 2015021484, and 201501C886. Model 57417 / Lot Numbers: 2014125438, 2015016317, 2015016695, 2015020566, 2015021401, 2015030166, 2015030435, 2015030629, and 201503C101. Model 57421 / Lot Numbers: 2015016410 and 2015017102. Model 96017 / Lot Numbers: 2015021260 and 2015030499 Model 96021 / Lot Number: 2015021134.
Other devices under the same FDA submission
1 device cleared under K875353 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DLP® 96428Medtronic, Inc. · CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS2 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page