Skip to content

SOMATOM Emotion 6 recallZ-1934-2015 · Siemens Medical Solutions USA, Inc · initiated Jun 4, 2015

Class II Terminated All recalls posted in June 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A potential safety issue involving two (2) broken screws and one (1) bowed, which is used to fix the Balance Weight of the DMS in the gantry, was found during a service call at the customer site.

FDA root cause category: Device Design

What the company is doing

A safety advisory notice, dated June 4, 2015, was sent to direct accounts explaining the problem and potential risk to patients and operators. Siemens engineers will begin replacing the bolts in May 2015.

Recall details

Recall number
Z-1934-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 4, 2015
Posted by the FDA
Jun 27, 2015
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
40 total
Product code
JAK
FDA submissions
K133424
Distribution
Nationwide Distribution.

Product description

SOMATOM Emotion 6; produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angels or spiral planes taken at different angles.

Code information (models, lots, serial numbers)

model numbers: 10165888 with serial numbers: 29161 29185 29186 29165 29176 29181 29198

Other devices under the same FDA submission

1 device cleared under K133424 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 71484).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page