Skip to content

Gentle Threads Interference Screw 9x35mm Full Thread recallZ-1945-2015 · Biomet, Inc. · initiated Jun 3, 2015

Class II Terminated All recalls posted in July 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Biomet, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Part Number: 905620 Lot: 824610 Gentle Threads Interference Screw 10x25mm Round Head is mixed with the Part Number: 905627 Lot: 753410 Gentle Threads Interference Screw 9x35mm Full Thread. If the incorrect screw is used various failures may occur which may ultimately lead to revision surgery or a delay in surgery.

FDA root cause category: Mixed-up of materials/components

What the company is doing

On 6/3/2015, URGENT MEDICAL DEVICE RECALL NOTICE notifications were sent to the affected distributors with instructions for locating and returning the affected product. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.

Recall details

Recall number
Z-1945-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 3, 2015
Posted by the FDA
Jul 1, 2015
Recalling firm
Biomet, Inc.
Quantity
41 units
Product code
HWC
FDA submissions
K982497
Distribution
Distribution-Domestic: None. International: Australia, Mexico, Chile, Costa Rica, Netherlands, and Brazil.

Product description

Gentle Threads Interference Screw 9x35mm Full Thread. An interference fixation screw for use in soft tissue reattachment procedures

Code information (models, lots, serial numbers)

Part Number: 905627 Lot: 753410

Other devices under the same FDA submission

6 devices cleared under K982497 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 71526).

Sources and updates