Slipped Capital Femoral Epiphysis (SCFE) Drivers of the… recallZ-1848-2015 · Pega Medical Inc. · initiated Jun 15, 2015

Class II Terminated All recalls posted in June 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Pega Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

SCFE Drivers of the Free-Gliding SCFE Screw System are recalled because the handle of the instruments may detach from the shaft if high torques are applied in counter-clockwise rotation.

FDA root cause category: Reprocessing Controls

What the company is doing

Pega Medical sent an Urgent Field Safety-Recall letter via email on June 15, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Advise on action to be taken by the customer: Identify the device with the same catalogue number and lot number Return the device to Pega Medical Inc. to the address shown below using the FedEx account. For further questions please call the firm at 450-688-5144 x. 242.

Recall details

Recall number
Z-1848-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 15, 2015
Posted by the FDA
Jun 26, 2015
Recalling firm
Pega Medical Inc.
Quantity
4 units
Product code
HWC
FDA submissions
K131591
Distribution
US Distribution to the states of : FL and TN.

Product description

Slipped Capital Femoral Epiphysis (SCFE) Drivers of the Free-Gliding SCFE Screw System with catalogue number SCF-MLD265 and SCF-MLD273.

Code information (models, lots, serial numbers)

Part Number: SCF-MLD265, LOT#: 130703-01. Part Number: SCF-MLD273, LOT#: 130703-02.