Contour Next Control Solution for Contour Next Blood… recallZ-2482-2015 · Bayer Healthcare, LLC · initiated Jun 22, 2015
Class II Terminated All recalls posted in August 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bayer Healthcare, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.
FDA root cause category: Device Design
What the company is doing
On 6/22/2015, Bayer Healthcare began sending URGENT - MEDICAL DEVICE FIELD CORRECTIVE ACTION notifications to affected Health Care Providers, Pharmacies, and Users. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers are instructed to always check the expiration, not to use and discard of expired control solutions. Also, only CONTOUR NEXT control solution should be used with the CONTOUR NEXT test strips. The control solution bottle should be shaken about 15 times before each use. The letter instructed customers to refer to their meter's user guide for further testing instructions. If test results outside the control range are received, another test should be performed following the instructions provided in the letter. If the problem persists, customers should contact Bayer Diabetes Support. Bayer will replace the control solution or provide new control ranges for the test strips, depending on the situation. The firm has implemented improvements to provide more consistent control solution product.
Recall details
- Recall number
- Z-2482-2015
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jun 22, 2015
- Posted by the FDA
- Aug 29, 2015
- Recalling firm
- Bayer Healthcare, LLC
- Quantity
- 5 million bottles of control solution
- Product code
- JJX
- FDA submissions
- K111268
- Distribution
- Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Slovenia, Spain, Sweden, Switzerland, UK, Albania, Azerbaijan, Bulgaria, Czech Republic, Estonia, Hungary, Korea, Latvia, Lithuania, Philippines, Romania, Serbia, Slovakia, and Thailand.
Product description
Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottle, 2.5mL. Indicated for use with fresh fingertip capillary whole blood samples.
Code information (models, lots, serial numbers)
SKU# 7314 (Level 2) and SKU# 7315 (Level 1) All Lots
Other devices under the same FDA submission
3 devices cleared under K111268 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- CONTOUR® NEXT EZ 9628Ascensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose145 reports
- CONTOUR® NEXT EZ 7252Ascensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose49 reports
- CONTOUR® NEXT EZ 7553Ascensia Diabetes Care US Inc. · Glucose Dehydrogenase, Glucose12 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page