Skip to content

i-Stat PT/INR Cartridges recallZ-2361-2015 · Abbott Point Of Care Inc. · initiated Jun 15, 2015

Class II Terminated All recalls posted in August 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain lots of PT/INR cartridges may generate a higher than typical number of PT/INR star outs potentially delaying the generation of patient results for PT/INR.

FDA root cause category: Under Investigation by firm

What the company is doing

Abbott Point of Care sent a recall letter dated June 15, 2015 via Fed Ex to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return the enclosed Business Reply Card (BRC). A Proof of Destruction From (FG-4533 in included with letter and is required to be returned to APOC as part of the credit process. Customers with questions were instructed to contact APOC Technical Support at 800-366-8020, Option 1 or their APOC representative.

Recall details

Recall number
Z-2361-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 15, 2015
Posted by the FDA
Aug 10, 2015
Recalling firm
Abbott Point Of Care Inc.
Quantity
423,168 cartridges
Product code
GJS
FDA submissions
K020355
Distribution
Worldwide Distribution - US (nationwide) and Internationally to China, Lebanon, Costa Rica, Singapore, Brazil, Jordan, Russia, Greece, Austria, India, Nigeria, United Kingdom, Korea, Saudi Arabia, Morocco, Slovenia, Portugal, Tunisia, Kazakhstn, Qatar, U.A.E. Israel, Taiwan, Mexico, Argentina, Poland, Italy, Hong Kong, Kuwait, Switzerland, Turkey, Ireland, Chile, Germany, and Australia.

Product description

i-Stat PT/INR Cartridges List number 03P89-24 The cartridges are packaged in sealed pouches, one per pouch. There are 24 cartridges packed in a box. a prothrombin time test useful in monitoring patients receiving oral anticoagulation therapy such as coumadin or warfarin.

Code information (models, lots, serial numbers)

All lots from S15007 to S15026 inclusive and All lots from T15037A to T15076A inclusive.

Other devices under the same FDA submission

1 device cleared under K020355 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.