Skip to content

ADVIA Chemistry Acetaminophen reagents recallZ-2416-2015 · Siemens Healthcare Diagnostics, Inc. · initiated Jun 30, 2015

Class II Terminated All recalls posted in August 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A change in the concentration of N-acetylcysteine (NAC) that may cause interference for the ADVIA Chemistry Systems Acetaminophen assay for certain lots.

FDA root cause category: Unknown/Undetermined by firm

What the company is doing

Siemens sent an Urgent Medical Device Correction letter dated June 30, 2015, to all affected customers.The letter identified the product the problem and the action needed to be taken by the customer. Siemens Healthcare Diagnostics is conducting a voluntary recall for the ADVIA Chemistry Acetaminophen reagent, for all in date and future kit lots. Customers were informed of the revised interference levels of N-acetyl cysteine (NAC) of 200 mg/dL, for the ADVIA Chemistry Acetaminophen assay. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within 30 days. Field service personnel will be sent an SKB (Service Knowledge Base) describing the issue and instructing them how to deal with customer questions. For further questions call 1 (800) 441-9250.

Recall details

Recall number
Z-2416-2015
Recall class
Class II
Status
Terminated
Initiated by the company
Jun 30, 2015
Posted by the FDA
Aug 19, 2015
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
1032
Product code
LDP
FDA submissions
K100200
Distribution
Worldwide Distribution - US (nationwide) and to the countries of : Canada, Netherlands, Great Britain, Norway, Sweden, Spain, Italy, Germany, France, Finland, Belgium, Portugal, Czech Republic, Israel, Poland, Turkey and Bahrain.

Product description

ADVIA Chemistry Acetaminophen reagents; for in vitro diagnostic use in the quantitative determination of acetaminophen in human serum and plasma (lithium heparin) on ADVIA Chemistry systems.

Code information (models, lots, serial numbers)

Catalog Number 07989138, Siemens Material Number (SMN) 10327381, Lot Numbers: 46251, 46564, 46369, 46370, 46619, 46782, 47001, 47157, 47558; with expiration dates June 30, 2015 through April 30, 2016.

Other devices under the same FDA submission

1 device cleared under K100200 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page