Skip to content

Stryker recallZ-2468-2015 · Stryker Sustainability Solutions · initiated Jul 10, 2015

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Sustainability Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Stryker Sustainability Solutions has received reports indicating that devices from one lot (lot# 3574704) were labeled as 100mm models (CB5LT) while the devices inside of the packaging were 75mm models (CB5ST).

FDA root cause category: Labeling mix-ups

What the company is doing

On 07/10/15 the firm sent out notification letters to all customers. The letter states for the customer to discontinue use of the affected lot, and to use other trocar lots reporcessed by the firm. The firm requests that the Recall Effectiveness Check Form be completed. The firm is going to be sending a sales representative to the facilities to check inventory for the affected trocars. A Recall Effectiveness Check will also be completed. If the form indicates that affected devices remain in inventory a prepaid shipping label will be issued for the return of the product. Customers will receive credit for all affected devices returned. The form must be completed even if no affected devices are found in inventory. Stryker Sustainability Solutions Complaint Hotline: 1-888-888-3433 x5555

Recall details

Recall number
Z-2468-2015
Status
Terminated
Initiated by the company
Jul 10, 2015
Posted by the FDA
Aug 27, 2015
Recalling firm
Stryker Sustainability Solutions
Quantity
90 devices
Product code
NLM
FDA submissions
K100080
Distribution
U.S. Nationwide Distribution to the following states of:: NV, NM, TX, and MI.

Product description

Stryker Sustainability Solutions CB5LT Ethicon Endo-Surgery, Inc. Excel Trocar Sleeve w/Stability Sleeve 5mm x 100mm Reprocessed by Stryker Sustainability Solutions Product Usage: The Reprocessed Ethicon ENDOPATH XCEL Universal Trocar Stability Sleeve is a sterile single patient use device with a radiolucent sleeve. The trocar sleeve contains two seals that accommodate instruments 5mm in diameter. Together, these two seals minimize gas leakage when instruments are inserted or withdrawn through the trocar. A stopcock valve is compatible with standard luer lock fitting and provides attachment for gas insufflation and desufflation. The stopcock is in the closed position when it is parallel to the sleeve.

Code information (models, lots, serial numbers)

Lot number 3574704 UDI Code: (01)00885825014421(17)180310(10)3574704 Model Number: CB5ST (labeled as CB5LT)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page