SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters.… recallZ-0003-2016 · Biosense Webster, Inc. · initiated Sep 10, 2015
Class II Terminated All recalls posted in October 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Biosense Webster, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
FDA root cause category: Device Design
What the company is doing
A Customer Notification Letter and Acknowledgement Form was distributed to customers on 09/10/15. One letter and form was sent via express mail, addressed to the Risk Management Office and one additional letter and form was sent by the firm's account representative to the Electrophysiology (EP/Cardiology Lab directly at the customer site. The firm's notification letter stated that they want to emphasize the following statement "The intra-cardiac ultrasound image will disappear if the CARTO 3 EP Navigation System power is disrupted and this may present a safety issue if the EP is using the ultrasound to monitor the patien during EP procedure. The ultrasound image will not reappear until the CARTO 3 System is restored." The firm will be updating the product labeling to further reinforce the precautionary statement. The firm requests that customers take the following actions: 1. Read the Field Safety Notification carefully. 2. Pass on this notification to anyone in the facility that needs to be informed of this issue, including appropriate clinical personnel involved in the use of SOUNDSTAR eco Catheters. 3. Review, complete, sign and return the attached Acknowledgement Form in accordance with the instructions on the form. 4. Maintain a copy of this letter with the product. 5. Maintain awareness of the Field Safety Notification For any questions the firm states to contact the BWI sales representative.
Recall details
- Recall number
- Z-0003-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 10, 2015
- Posted by the FDA
- Oct 1, 2015
- Recalling firm
- Biosense Webster, Inc.
- Quantity
- 1035 (U.S.)
- Product code
- OBJ
- FDA submissions
- K140318
- Distribution
- Worldwide Distribution. US nationwide, Italy, Canada, Japan, China, Singapore, Colombia, England, Russia, Spain, France, Germany, Finland, Ireland, Australia, Sweden, Hungary, India, Brazil, Slovenia, Netherlands, Mexico, and South Africa.
Product description
SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. For intra-cardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart.
Code information (models, lots, serial numbers)
Catalog #: 10439236, 10439072, 10439011, 10438577
Other devices under the same FDA submission
2 devices cleared under K140318 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- SOUNDSTAR 10439236Biosense Webster Inc · Catheter, ultrasound, intravascular4 deaths107 reports
- SOUNDSTAR 10439011Biosense Webster Inc · Catheter, ultrasound, intravascular2 deaths90 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page