Samsung GM60A-40S recallZ-0263-2016 · NeuroLogica Corporation · initiated Sep 14, 2015

Class II Terminated All recalls posted in November 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from NeuroLogica Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When the system is driving over a non-conductive floor and then over a conductive, grounded part, proper control of the drive behavior may not be possible. The system can be stopped by releasing the drive handle or by pressing the emergency switch. Proper behavior is re-enabled when the system is rebooted.

FDA root cause category: Device Design

What the company is doing

An Urgent - Field Safety Notice identifying the affected products, problem involved, root cause, and potential hazards, was sent to customers. In the event the issue occurs, users can release the drive handle or press the emergency switch and the unit will stop. Then, reboot the system. If unintended movements occur, please contact local representative. To correct the issue, service engineers will visit to apply the corrective action. For further information or support concerning this problem, contact local representative.

Recall details

Recall number
Z-0263-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Sep 14, 2015
Posted by the FDA
Nov 10, 2015
Recalling firm
NeuroLogica Corporation
Quantity
7 devices
Product code
IZI
FDA submissions
K142492, K150097
Distribution
Distributed to the states of NC, OH, MA, and MO.

Product description

Samsung GM60A-40S Intended for use in generating radiographic images of human anatomy.

Code information (models, lots, serial numbers)

Serial Numbers: 50EAM9BG200004X, 50EAM9BG200001A 50EAM9BG200002P, 50EAM9BG200003D, 50EAM9BG200005N, 50EAM9BG200006E,  50EAM9BG200007F

Other devices under the same FDA submission

2 devices cleared under K142492, K150097 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.