Palindrome Precision Chronic Catheter Kit Symmetrical Tip… recallZ-0331-2016 · Covidien LLC · initiated Oct 14, 2015
Class II Terminated All recalls posted in November 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Covidien LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Incorrectly packaged with a 90 degree bend at the distal tip
FDA root cause category: Packaging process control
What the company is doing
The firm, Medtronic (Covidien), sent an "URGENT MEDICAL DEVICE RECALL" letter dated 10/16/2015 to their customers. The letter described the product, problem, and actions to be taken. Customers were instructed to quarantine and discontinue use of the product, return affected product and complete and return the RECALLED PRODUCT RETURN FORM via Fax to Medtronic to (800)-895-6140 or email to: [email protected]/Distributor to (203) 492-7719 or email to: [email protected]. Questions or concerns contact your Medtronic representative or Customer Service at (800) 882-5878.
Recall details
- Recall number
- Z-0331-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 14, 2015
- Posted by the FDA
- Nov 24, 2015
- Recalling firm
- Covidien LLC
- Quantity
- 365 units
- Product code
- NYU
- FDA submissions
- K123196
- Distribution
- Worldwide distribution: US (nationwide) and countries of: Canada, Belgium, Chile, Panama, Denmark, France, Germany , Netherlands,Spain, Switzerland, and United Kingdom.
Product description
Palindrome Precision Chronic Catheter Kit Symmetrical Tip, Tal VenaTrac" Stylet 14.5 Fr/Ch (4.8mm) X 33 cm Item Number: 8888145042P The Palindrome chronic catheter is intended for acute and chronic hemodialysis, apheresis, and infusion. It may be inserted either percutaneously or by cutdown. Catheters grater than 40 cm implant length are indicated for femoral insertion.
Code information (models, lots, serial numbers)
Lot Numbers: 1504700102, 1516900090, 1425000091
Other devices under the same FDA submission
70 devices cleared under K123196 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Palindrome 8888119369PMozarc Medical US LLC · Catheter, hemodialysis, implanted3 recalls1 death14 reports
- Palindrome 8888145044CPMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 deaths30 reports
- Palindrome 8888119364PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123404PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death17 reports
- Palindrome 8888123409PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls1 death12 reports
- Palindrome 8888145014PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall62 reports
- Palindrome 8888123410PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls9 reports
- Palindrome 8888145040PMozarc Medical US LLC · Catheter, hemodialysis, implanted1 recall1 death58 reports
- Palindrome 8888123400PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls5 reports
- Palindrome 8888128454PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls5 reports
- Palindrome 8888133504PMozarc Medical US LLC · Catheter, hemodialysis, implanted2 recalls3 reports
- Palindrome 8888119365PMozarc Medical US LLC · Catheter, hemodialysis, implanted, coated2 recalls2 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 72416).
- Z-0329-2016 · TerminatedPalindrome" Chronic Catheter Kit Symmetrical Tip, Tal VenaTrac" Stylet 14.5 Fr/Ch (4.8mm) X 33 cm Item Number: 8888145042 The Palindrome chronic catheter is intended for acute an…
- Z-0330-2016 · TerminatedPalindrome Precision Chronic Catheter Kit Symmetrical Tip 14.5 Fr/Ch (4.8mm) X 33 cm Item Number: 8888145017P The Palindrome chronic catheter is intended for acute and chronic he…
- Z-0332-2016 · TerminatedPalindrome Precision Chronic Catheter Kit Symmetrical Tip, Tal VenaTrac" Stylet 14.5 Fr/Ch (4.8mm) X 33 cm Item Number: 8888145046C The Palindrome chronic catheter is intended for…
- Z-0333-2016 · TerminatedPalindrome Precision Chronic Catheter Kit Symmetrical Tip, Pre-Curved Shaft and Tal VenaTrac" Stylet 14.5 Fr/Ch (4.8mm) X 33 cm Item Number: 8888145061P The Palindrome chronic cat…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page