Skip to content

Synthes Application Instrument for Sternal ZIPFIX.… recallZ-0276-2016 · Synthes (USA) Products LLC · initiated Oct 19, 2015

Class II Terminated All recalls posted in November 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Synthes (USA) Products LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Certain lots of the Application Instrument for Sternal ZIPFIX that were previously recalled on August 19, 2015 and subsequently repaired and returned to customers between August 26, 2015 and September 23, 2015. The screws on the repaired instruments have the potential to break or loosen. DePuy Synthes is requesting the return of the repaired device in exchange for a new device.

FDA root cause category: Device Design

What the company is doing

Communication to the accounts having record of receiving this product was sent on October 19, 2015.

Recall details

Recall number
Z-0276-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 19, 2015
Posted by the FDA
Nov 14, 2015
Recalling firm
Synthes (USA) Products LLC
Quantity
267
Product code
JDQ
FDA submissions
K110789
Distribution
Distributed in the states of KY, UT, CA, WI, OH, CO, IL, TX, OK, WA, MI, LA, MA, PA and NE.

Product description

Synthes Application Instrument for Sternal ZIPFIX. Indications for use include primary or secondary closure/repair of the sternum following sternotomy or fracture of the sternum to stabilize the sternum and promote fusion.

Code information (models, lots, serial numbers)

Synthes Application Instrument for Sternal ZIPFIX, Part Number 03.501.080. Lot Numbers 3783913, 7516728, 7606881, 7659168, 7666085, 7671934, 7689244, 7694377, 7720599, 7738572, 7738573, 7740498, 7767497, 7803768, 7806881, 7818677, 7818682, 7821672, 7827088, 7831855, 7833606, 8068078, 8130898, 8145793, 8207769, and 8215969.

Other devices under the same FDA submission

4 devices cleared under K110789 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page