Amphirion Deep OTW PTA Balloon Catheter recallZ-0243-2016 · Medtronic Vascular, Inc. · initiated Oct 23, 2015
Class II Terminated All recalls posted in November 2015
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Vascular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Medtronic has discovered that select lots of the Amphirion Deep PTA Balloon Dilation Catheter have a potential breach in the pouch seal enclosing the sterilized product.
FDA root cause category: Equipment maintenance
What the company is doing
Beginning October 23, 2015, Medtronic will send the customer letter and confirmation form to all identified US customers via 2-day UPS mail delivery. Internationally, communications will be sent per locally approved methods.
Recall details
- Recall number
- Z-0243-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 23, 2015
- Posted by the FDA
- Nov 6, 2015
- Recalling firm
- Medtronic Vascular, Inc.
- Quantity
- 1652 units in US; 4243 units OUS.
- Product code
- DQY
- FDA submissions
- K042624, K050073, K052791, K083919
- Distribution
- Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia Austria Belgium Brazil China Colombia Costa Rica Czech Republic Denmark Dominican Republic Egypt France Georgia Germany Hong Kong Hungary India Iran Ireland Israel Italy Japan Jordan Kazakhstan Korea Kuwait Latvia Lebanon Lithuania Luxembourg Mexico Netherlands Norway Oman Pakistan Panama Peru Philippines Poland Portugal Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Syria Taiwan Thailand Trinidad and Tobago Turkey United Arab Emirates United Kingdom and Uruguay. .
Product description
Amphirion Deep OTW PTA Balloon Catheter: Model numbers: US: AMP015020152 AMP020150152 AMP025120152 AMP030080152 AMP040040152 AMP225210152 AMP020040152 AMP025040152 AMP025150152 AMP030120152 AMP040080152 AMP253210152 AMP020080152 AMP025080002 AMP030040002 AMP030150152 AMP040120152 AMP335210152 AMP020120152 AMP025080152 AMP030040152 AMP035120152 AMP040150152 AMP354210152; OUS: AMD025150152 AMD030150152 AMD040120002 AMD335210002 AMD030040002 AMD035040152 AMD040120152 AMD335210152 AMD030040152 AMD035120002 AMD040150002 AMD354210002 AMD030080002 AMD035150002 AMD040150152 AMD354210152 AMD030080152 AMD035150152 AMD225210002 AMD030120002 AMD040040152 AMD225210152 AMD030120152 AMD040080002 AMD253210002 AMD030150002 AMD040080152 AMD253210152 Cardiovascular: Intended for use in Angioplasty
Code information (models, lots, serial numbers)
Show the full code information (6k characters)
Model number/lot numbers: AMD015020002: 209633629; 209882401; 209981044; AMD015020151; 209436628; AMD015020152; 209821677; 209900858; 210046728 AMD020040002; 209425805; 209601495; 209760077; 209821680; AMD020040152; 209406894;209425802;209471311; 209532785; 209551995;209601496; 209664056; 209759618; 209759619; 209855564; AMD020080002: 209821676; 209855251; 209901400; 210005917; 210046729; AMD020080152: 209436625; 209470469; 209470470; 209532768; 209551997 209760064; 209760065; 209855561; 209900855; AMD020120002; 209722681; 209760093 AMD020120152: 209470465; 209470468; 209532777; 209551996; 209601499 209722674; 209760062; 209760063; 209821673; 209855560; 209900846; 209911605; AMD020150002: 209532784; 209601620; 209664054; 209722667; 209760060; 209981040; 210005913; AMD020150152: 209425766; 209436575; 209509786; 209532790; 209633607; 209664057; 209722685; 209760082; 209855565; 209882402; 209900852; 209911606; 210046730; AMD025040152: 209471307; 209722670; 209821666; 209882403; 209900853; 210046726; AMD025080152; 209406892; 209425768; 209471303; 209509782; 209509783; 209532791; 209601621; 209633611; 209664058; 209760080; 209821679; 209855566; 209882404; 209911600; AMD025120002: 209532776; 209551998; 209760088; AMD025120152: 209470467;209532762; 209664055; 209760069; 209794025; 209882405; 209900851; 209917255; AMD025150002: 209425767; 209553428; 209633606; 209722660; 209760091; 209911603; AMD025150152: 209425771; 209436620; 209470477; 209509781; 209553429; 209633608; 209722663; 209760076; 209900857; 209911517; 210005909; 210046732; AMD030040002: 209821681; AMD030040152: 209471306: 209509785: 209553420; 209601625; 209760061; AMD030080002: 209436579; 209855252; AMD030080152: 209470478; 209532782; 209553421; 209722662; 209759614: 209759615; 209917248; 209981041; 210005915; AMD030120002: 209425806; 209509784; 209601627;209760094 AMD030120152: 209601628; 209722477; 209760068; 209917242; 209981042. AMD030150002: 209530579; 209553422; 209759610; 209759611; 209821674 AMD030150152: 209507756; 209507757; 209633602; 209722678; 209759612; 209759613; 209855257; AMD035040152: 209634197; 209980257; AMD035120002: 209436577; 209553423; AMD035150002: 209633627; AMD035150152: 209436574; 209633625; 209635221; 209980247; 210046727; AMD040040152: 209425770; 209470473; 209532789; 209553424; 209601629; 209855563; 209900854; AMD040080002: 209436627; AMD040080152: 209601630; 209634198; 209882406; 209980249; 210005903; 210046733; AMD040120002: 209425807; AMD040120152: 209532778; 209601631; 209635220; 209855562; 209911519; 210046725; AMD040150002: 209470464; AMD040150152: 209425763; 209471313; 209532761; 209553425; 209823991; AMD225210002: 209601633; 209722478; 209760078; 209917256; 209980250; 210005904; AMD225210152: 209436626; 209471308; 209532760; 209664048; 209722661; 209760071; 209760072; 209821675; 209855568; 209855569; 209917244; 209917245; 209917246; 209980252; 209980253; 209980254; AMD253210002: 209425769; 209436621; 209532788; 209553426; 209633600; 209722680; 209760070; 209968089; 209981043; 210005912; AMD253210152: 209471301; 209507758; 209509780; 209532771; 209553427; 209633604; 209664051; 209760066; 209760067; 209794020; 209821678; 209882407; 209882408; 209968083; 209968088; 209968091; 209980255; 209980256; 210005914; AMD335210002: 209633603; 209664050; 209760087; 209855567; 209882409; 209968085; AMD335210152; 209470466; 209507754; 209507755; 209633601; 209664049 209760079; 209793539; 209821554; 209855259; 209882410; 209968084; 209968092; 209980259; AMD354210002: 209470475; AMD354210152: 209470474; 209532787; 209633610; 209722475; 209980248; 210005905; AMP015020152: 209471310: 209532780; 209633623; 209722671; 209760084; 209794023; 209900845; AMP020040152: 209436622; 209532772; 209633618; 209722666; 209760074; 209821557; 209900859; AMP020080152: 209425773; 209471314; 209532786; 209633621; 209722677 209821558; 209900850; 209917250; 210006063; AMP020120152: 209470476; 209532765; 209722683; 209821664; 209855254; AMP020150152: 209436576; 209471312; 209532781; 209633620; 209722673; 209760090; 209821672; 209917249; AMP025040152: 209633628; 209760083; 209821667; 209917253 ; AMP025080002: 209633624; AMP025080152: 209471304; 209532774; 209633630; 209821668; 209917247; AMP025120152: 209470472; 209532766; 209633612; 209722679; 209794021; 209917252; 209980258; 210005907; AMP025150152: 209471302; 209722474; 209794022; 210046785; AMP030040002: 209722672; AMP030040152: 209436578; 209532779; 209633613; 209722668; 209760095; 209821665; 209900856; AMP030080152: 209532769; 209633619; 209722664; 209760089; 209821671; 209917254; 210005919; AMP030120152 : 209471309; 209532773; 209633617; 209722665; 209760086; 209821670; 209882411; AMP030150152: 209436623; 209532764; 209633615; 209722479; 209760085; 209794024; AMP035120152: 209470479; 209633616; 209722676: AMP040040152; 209470471; 209532767; 209633622; 209760081; 209821555; 209882412; 209900844; 209917251; 210006061; AMP040080152: 209821559; 210046724; AMP040120152: 209471315; 209855253; 209882413; 210005918; AMP040150152: 209532775; 209722682; 209821669; 209900847; 210005911; AMP225210152: 209436624; 209471305; 209532763; 209759616; 209759617; 209821660; 209821661; 209821662; 209821663; 209855255; 209855256; 209968082; 209968087; 209968090; 209980251; 210005906; AMP253210152; 209425778; 209471300; 209532770; 209633614; 209664053; 209722675; 209760073; 209821556; 209900849; 210006060; AMP335210152: 209633605; 209722669; 209760096; 209917243; AMP354210152: 209532783; 209760092; 209900848.
Other devices under the same FDA submission
1 device cleared under K042624, K050073, K052791, K083919 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- AMPHIRION DEEP AMP020150152Medtronic, Inc. · CATHETER, PERCUTANEOUS1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.