rainbow Reusable Sensors. RAINBOW DCI-DC3… recallZ-0338-2016 · Masimo Corporation · initiated Oct 28, 2015

Class II Terminated All recalls posted in November 2015

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Masimo Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

These sensors were manufactured with incompatible configurations. This could result in sensors that will provide either no readings or inaccurate readings for the SpCO and SpMet parameters.

FDA root cause category: Finished device change control

What the company is doing

Masimo sent an Urgent Product Recall letter on October 28, 2015, to all affected customers to inform them that Masimo has identified specific lots of sensors that were manufactured with incompatiable configurations. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete the tracking/verification form and email to Masimo. Customers with questions were instructed to contact Masimo Technical Services at http://Service.Masimo.com. For questions regarding this recall call 949-297-7000.

Recall details

Recall number
Z-0338-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 28, 2015
Posted by the FDA
Nov 25, 2015
Recalling firm
Masimo Corporation
Quantity
3476 units total (3322 units in US)
Product code
DQA
FDA submissions
K080238
Distribution
Worldwide Distribution - US (nationwide) and Internationally to Australia, Costa Rica, France, Germany, Kuwait, Martinique, Netherlands, Norway, South Korea, and Turkey.

Product description

rainbow Reusable Sensors. RAINBOW DCI-DC3, SpO2/SpCO/SpMet, Part No. 2201; RAINBOW DCI-dc8, SpO2/SpCO/SpMet, Part No. 2407; RAINBOW DCI, SpO2/SpCO/SpMet, Part No. 2696; RAINBOW DCIP, PED, SpO2/SpCO/SpMet, Part No. 2697. rainbow reusable sensors are indicated for spot check or continuous non-invasive monitoring of arterial saturation (SpO2), arterial carboxyhemoglobin saturation (SpCO), arterial methemoglobin (SpMet), and pulse rate.

Code information (models, lots, serial numbers)

2201 2407 2696 2697

Other devices under the same FDA submission

11 devices cleared under K080238 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.