Accu-Chek Inform II Base Unit as a part of the Accu-Chek… recallZ-1172-2016 · Roche Diagnostics Operations, Inc. · initiated Feb 11, 2016
Class II Terminated All recalls posted in March 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Roche Diagnostics Operations, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Accu-Chek Inform II Base Unit might produce physical transmission errors in the form of data loss in the communication between the meter and the Data Management Systems (DMS). The issue can lead to the data loss or in the worst case to an erroneous assignment of the patient data (patient mismatch). The issue will only occur at sites using POTCT1-A communication via USB.
FDA root cause category: Software Design Change
What the company is doing
Roche sent an URGENT MEDICAL DEVICE RECALL CORRECTION letter dated February 11, 2016, via UPS Ground (receipt required) to the consignee. The Consignee instructions were as follows: If currently using USB connectivity, follow the instructions outlined in this Urgent Medical Device Correction (UMDC) to determine if your base unit is affected by the issue. Until the new software is available, do not use USB connectivity if the base unit is affected by this issue; switch the connectivity method to Ethernet (TCP/IP) or Wi-Fi. Roche is not requesting return of affected product. Any returned product will be managed according to current procedures. Consignees with questions were instructed to contact the ACCU-CHEK Customer Care Service Center at 1-800-440-3638. For questions regarding this recall call 317-521-3911.
Recall details
- Recall number
- Z-1172-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Feb 11, 2016
- Posted by the FDA
- Mar 14, 2016
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Quantity
- US 5,604 devices, OUS 91,925
- Product code
- LFR
- FDA submissions
- K121679
- Distribution
- Nationwide Distribution including AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI, and WV.
Product description
Accu-Chek Inform II Base Unit as a part of the Accu-Chek Inform II Blood Glucose Monitoring System Accu-Chek inform II Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings. This system should only be use with single-use, auto-disabling lancing devices.
Code information (models, lots, serial numbers)
Model/Catalog/Part Number: 05060290001; serial numbers: UU41000000 through UU41064963 and UU41064964 through UU41122741.
Other devices under the same FDA submission
4 devices cleared under K121679 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Accu-Chek Inform II Instrument with RF card 5060303001Roche Diagnostics GmbH · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER46 reports
- Accu-Chek Inform II Base Unit 07671717190Roche Diagnostics GmbH · System, Test, Blood Glucose, Over The Counter6 reports
- Accu-Chek Inform II Instrument with RF card 04882458001Roche Diagnostics GmbH · System, Test, Blood Glucose, Over The Counter4 reports
- Accu-Chek Inform II Base Unit Light 08376824190Roche Diagnostics GmbH · SYSTEM, TEST, BLOOD GLUCOSE, OVER THE COUNTER1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page