Skip to content

ClosureFast Micro Introducer Sheath Set 7Fr/Ch(2.3mm) x 7… recallZ-1029-2016 · Argon Medical Devices, Inc · initiated Feb 17, 2016

Class II Terminated All recalls posted in March 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Argon Medical Devices, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The incorrect sheath was assembled to the device so that the size 7 products received 11 cm sheaths and the size 11 product received 7 cm sheaths.

FDA root cause category: Process control

What the company is doing

On February 17, a letter was sent to Medtronic notifying them of a voluntary recall.

Recall details

Recall number
Z-1029-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 17, 2016
Posted by the FDA
Mar 1, 2016
Recalling firm
Argon Medical Devices, Inc
Quantity
2000 units
Product code
DRE
FDA submissions
K780126
Distribution
Distributed in IL.

Product description

ClosureFast Micro Introducer Sheath Set 7Fr/Ch(2.3mm) x 7 cm REF MIS-F707 and ClosureFast Introducer Sheath Set 7Fr/Ch(2.3mm) x 11 cm REF IS-7F11. Indicated for use in arterial and venous procedures requiring percutaneous introduction of intravascular devices.

Code information (models, lots, serial numbers)

Lot 1125585 (MIS-F707) Lot 11126021 (IS-7F11)

Other devices under the same FDA submission

16 devices cleared under K780126 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page