Simplexa HSV 1&2 Direct (MOL2150) with Direct Amplification… recallZ-1272-2016 · Focus Diagnostics Inc · initiated Feb 10, 2016
Class I Terminated All recalls posted in April 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Focus Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Focus Diagnostics is recalling the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850) because it may result in a false negative result, a false positive result or Error code (s) 500 or 505, and it may result in higher invalid rates.
FDA root cause category: Under Investigation by firm
What the company is doing
A customer notification letter dated 2/10/16 was sent to all their customers who received the Direct Amplification Discs (DAD) supplied with the Simplexa HSV 1 & 2 Direct (MOL2150), Simplexa Flu A/B & RSV Direct (MOL2650) and Simplexa Group A Strep Direct kits (MOL2850). The letter informs the customers that Focus Diagnostics is recalling MOL2150 and MOL2650 kits because it may result in a false negative result, a false positive result or Error Code(s) 500 or 505. Additionally, some customers are reporting insufficient specimen volume errors which may result in higher invalid rates. In those cases, the error results in a Report Note Statement of Insufficient Specimen Volume. The letter informs the customers of the recommendations to be taken for the identified kits and the risks involved for positive results, negative results, and delayed results. The letter informs the customers of the actions to be taken. Customers are instructed to sign the achnowledgement form and email the form to [email protected] or fax back to Focus Diagnostics Technical Services at (562) 240-6526 within 10 business days. On 2/11/16, Focus Diagnostics sent to their customers an amendment to the customer notification letter to clarify what is considered an "early Ct", Focus would consider Internal Control (IC) Ct's less than 25 as "early Ct". Customers are informed that if an internal control Ct is less than 25, additional testing may be necessary. On 2/15/16, Focus Diagnostics sent to their customers an amendment to the customer notification letter to inform them to determine if they had instances of early Cts less than 25, then to please refer to Section III Analyzing Runs, pages 6-10 and 6-15 in the Focus Diagnostics, Operator Manual for Integrated Cycler Studio 6.0 For Use with IVD Assays in the USA. Also, for a more detailed analysis, customers are instructed to view excerpts from the ICS Operator Manual covering Section III, pages 6-10 and 6-16. On 3/17/16,
Recall details
- Recall number
- Z-1272-2016
- Recall class
- Class I (most serious)
- Status
- Terminated
- Initiated by the company
- Feb 10, 2016
- Posted by the FDA
- Apr 2, 2016
- Recalling firm
- Focus Diagnostics Inc
- Quantity
- 1845 units
- Product code
- PGH
- FDA submissions
- K141458, K150962
- Distribution
- Worldwide Distribution - US Nationwide in the states of: TX, CA, MA, NY, WA, MI, ME, IN, NJ, WV, OH, PA, HI, NH, CO, MN, FL, SC, IN, IA, TN, OR, AZ, WI, TN, IL. and the countries of: Australia, Brazil, Canada, Dominican Republic, Denmark, Ecuador, France, Germany, Israel, Kuwait, Spain, Sweden, Slovenia, Thailand, and United Kingdom.
Product description
Simplexa HSV 1&2 Direct (MOL2150) with Direct Amplification Disc (MOL1451, MOL1452, MOL1455) Model Number: MOL2150, MOL1451, MOL1452, MOL1455
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Item Number MOL1451, Lot/Serial: 151870, 152535,152359, 152550, 152551, 152756, 152770, 160008, 160009, 160198, 160201, 160248, 160262. Item Number MOL1455, Lot/Serial: 29230, 29232, 29233, 29453, 29454, 29669, 29670, 29671, 29689, 29690, 29691, 29692, 29845, 29846, 29847, 29848, 29849, 29850, 30271, 30272, 30273, 30274, 30275, 30609, 30610, 30611, 30612, 30613, 30615, 30616, 30617, 30619, 30620. 3M Manufacturing Lot# (Pouch and White Box of 24 DAD): 2140887, 2140332, 2127423, 2159531, 2154010, 2165240, 2173319, 2169601, 2176090, 2181518, 2181519, 2198924, 2163486, 2171337, 2176091, 2165236, 2183651, 2185129, 2216981, 2213066, 2210371, 2208204, 2205280, 2255456, 2255456, 2225778, 2230958, 2233449, 2265481 2258543.
Other devices under the same FDA submission
1 device cleared under K141458, K150962 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Simplexa™ HSV 1 & 2 Direct MOL2150Diasorin Molecular LLC · Hsv-1 And Hsv-2 Cns Nucleic-Acid Based Panel1 death1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page