The Livongo Health In Touch Blood Glucose Monitoring System… recallZ-1587-2016 · Livongo Health · initiated Feb 18, 2016

Class II Terminated All recalls posted in April 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Livongo Health or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The instructional insert for the In Touch Control Solution for use with the In Touch meter and In Touch strips as a quality control check did not contain the limitation of sensor accuracy: (1) with the use of acetaminophen (Tylenol) (2) with the use of Ascorbic Acid (vitamin C) supplementation (3) in the presence of uric acid

FDA root cause category: Error in labeling

What the company is doing

The recall was conducted in three distinct events. Members were first sent an e-mail of the Recall Notification - Livongo Test Strip Insert (Event 1) containing information regarding the test strip insert discrepancy. Following the e-mail distribution, a hard copy of the same information was distributed (Event 2) to members via United States Postal\ Service (USPS) mail. Lastly, a second e-mail (Event 3) containing the same information was sent to members. The distribution list for the second (hardcopy distribution) and third (e-mail distribution) events was a subset of the initial member distribution.

Recall details

Recall number
Z-1587-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Feb 18, 2016
Posted by the FDA
Apr 29, 2016
Recalling firm
Livongo Health
Quantity
63,000 units
Product code
NBW
FDA submissions
K133584
Distribution
US consignees: Arizona, California, Delaware, Florida, Georgia, Illinois, Kentucky, Maryland, Michigan, Minnesota, Mississippi, Missouri, Montana, New Jersey, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia and Washington.

Product description

The Livongo Health In Touch Blood Glucose Monitoring System (BGMS), is an Over-The-Counter (OTC) system designed for the self- monitoring of blood glucose by persons with diabetes in home settings to aid in their diabetes management.

Code information (models, lots, serial numbers)

Model/Reorder Number 3614-001-001 with the following lot numbers 890071 890072, 890073, 890074, 890075, 890076, 890077, 890078 and 890079.

Other devices under the same FDA submission

2 devices cleared under K133584 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.