Skip to content

TOPAZ MicroDebrider 0.8 mm REF AC4040-01 recallZ-1321-2016 · ArthroCare Corporation · initiated Mar 3, 2016

Class II Terminated All recalls posted in April 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ArthroCare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During functional testing of devices after real time aging, it was determined that product failed to meet shelf-life at the expiration date indicated on the product labels.

FDA root cause category: Package design/selection

What the company is doing

Consignees were notified via letter on 3/3/2016. Consignees were instructed to quarantine the recalled products and complete an Inventory Return Certification Form.

Recall details

Recall number
Z-1321-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 3, 2016
Posted by the FDA
Apr 5, 2016
Recalling firm
ArthroCare Corporation
Quantity
26,903 devices
Product code
GEI
FDA submissions
K080282
Distribution
Worldwide Distribution-US (nationwide) and the following countries-Austria, Belgium, Bulgaria, Canada, Denmark, Estonia, Finland, France, Germany, Gibraltar, Great Britain, Greece, Ireland, Israel, Italy, Latvia, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, China, Hong Kong, India, Lebanon, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arabic Emirates, Vietnam, Australia, Japan, New Zealand, Brazil, and Chile.

Product description

TOPAZ MicroDebrider 0.8 mm REF AC4040-01; TOPAZ MicroDebrider IFS 0.8 mm REF ACH4040-01; TOPAZ MicroDebrider 0.8 mm REF Q6000-01; TOPAZ EPF MicroDebrider 0.8 mm 45 REF Q6002-01

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

1031506, 1031507, 1032123, 1032325, 1032741, 1032742, 1033443, 1033965, 1033966, 1034397, 1039120, 1041765, 1044089, 1044090, 1045862, 1048237, 1048933, 1049185, 1052071, 1056212, 1060586, 1060979, 1063394, 1064599, 1067696, 1067748, 1069174, 1071358, 1074125, 1075714, 1070521, 1070812, 1071974, 1078025, 1077499, 1082811, 1084095, 1084094, 1085458, 1086436, 1086437, 1087917, 1094313, 1097573, 1103824, 1107183, 1107297, 1113562, 1113563, 1114534, 1114535, 1114536, 1115896, 1031518, 1032308, 1032309, 1032310, 1032755, 1032756, 1033972, 1033973, 1034225, 1034226, 1035283, 1035284, 1038547, 1043507, 1046556, 1047064, 1048722, 1049266, 1051812, 1052070, 1054443, 1055730, 1059865, 1063661, 1067691, 1067273, 1067747, 1068019, 1069172, 1071357, 1072853, 1072854, 1070497, 1070808, 1071971, 1076849, 1077450, 1079140, 1079220, 1078115, 1081753, 1082361, 1082810, 1086380, 1094312, 1094312, 1094721, 1097571, 1101041, 1107295, 1109045, 1111935, 1111936, 1113534, 1113535, 1113536, 1114532, 1115888, 1115891, 1116959, 1116130, 1116720, 1117209, 1117214, 1031341, 1032311, 1033761, 1034446, 1035691, 1035692, 1041775, 1043962, 1046557, 1048236, 1050067, 1052651, 1057008, 1058808, 1059867, 1063665, 1064594, 1067695, 1068021, 1071319, 1071320, 1073441, 1074876, 1077498, 1080520, 1114533, 1115893, 1031340, 1032962, 1033266, 1045359, 1047065, 1051331, 1056849, 1064592, 1067694, 1070709

Other devices under the same FDA submission

3 devices cleared under K080282 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page