MEVION S250 for Proton Radiation Therapy recallZ-1529-2016 · Mevion Medical Systems, Inc. · initiated Mar 15, 2016

Class II Terminated All recalls posted in April 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death.

FDA root cause category: Device Design

What the company is doing

Customer notification letters were sent to customers on March 15, 2016, by certified mail, return receipt requested, as well as by email. The letter explains the problem, description of potential harm, and the cause containment, and mitigation. The cause of the malfunctioning sensors was discovered to be a non-conforming limit switch assembly. A Mevion Field Service Engineer will immediately inspect all such parts in the field for similar non-conformance, and replace those parts if necessary. In addition, a new periodic preventive maintenance check will be performed to individually test each of the redundant sensors at all sites. In the interim period, on a daily basis, until the inspection, repair, and the preventive maintenance check is completed at any site, Mevion Field Service will check that the Applicator Door interlock sensors are working properly. Mevion will also update the inspection procedure for incoming parts to ensure that any non-conformance is identified early in the manufacturing process.

Recall details

Recall number
Z-1529-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 15, 2016
Posted by the FDA
Apr 20, 2016
Recalling firm
Mevion Medical Systems, Inc.
Quantity
5
Product code
LHN
FDA submissions
K120676
Distribution
US nationwide distribution to MO, OK, FL, and NJ.

Product description

MEVION S250 for Proton Radiation Therapy

Code information (models, lots, serial numbers)

S250-0002

Other devices under the same FDA submission

1 device cleared under K120676 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.