MEVION S250 for Proton Radiation Therapy recallZ-1529-2016 · Mevion Medical Systems, Inc. · initiated Mar 15, 2016
Class II Terminated All recalls posted in April 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Possibility of Gantry Motion with Aperture or Compensator only partially inserted. An Aperture may fall out during Gantry motion. It is possible that it can hit a patient and cause traumatic injury or death.
FDA root cause category: Device Design
What the company is doing
Customer notification letters were sent to customers on March 15, 2016, by certified mail, return receipt requested, as well as by email. The letter explains the problem, description of potential harm, and the cause containment, and mitigation. The cause of the malfunctioning sensors was discovered to be a non-conforming limit switch assembly. A Mevion Field Service Engineer will immediately inspect all such parts in the field for similar non-conformance, and replace those parts if necessary. In addition, a new periodic preventive maintenance check will be performed to individually test each of the redundant sensors at all sites. In the interim period, on a daily basis, until the inspection, repair, and the preventive maintenance check is completed at any site, Mevion Field Service will check that the Applicator Door interlock sensors are working properly. Mevion will also update the inspection procedure for incoming parts to ensure that any non-conformance is identified early in the manufacturing process.
Recall details
- Recall number
- Z-1529-2016
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Mar 15, 2016
- Posted by the FDA
- Apr 20, 2016
- Recalling firm
- Mevion Medical Systems, Inc.
- Quantity
- 5
- Product code
- LHN
- FDA submissions
- K120676
- Distribution
- US nationwide distribution to MO, OK, FL, and NJ.
Product description
MEVION S250 for Proton Radiation Therapy
Code information (models, lots, serial numbers)
S250-0002
Other devices under the same FDA submission
1 device cleared under K120676 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Mevion Medical Systems S250Mevion Medical Systems, Inc. · System, Radiation Therapy, Charged-Particle, Medical4 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.