Skip to content

TC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE… recallZ-1537-2016 · Smith & Nephew, Inc. · initiated Mar 11, 2016

Class II Terminated All recalls posted in April 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Smith & Nephew, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Affected Devices may contain a dimensional deviation that could potentially result in the surgeon being unable to seat the polyethylene (PE) Insert on the affected VKS tibial base plate during surgery.

FDA root cause category: Process control

What the company is doing

The firm, Smith & Nephew, issued an "URGENT FIELD SAFETY NOTICE: Medical Device Field Safety Corrective Action/Recall" letter dated 3/10/2016 to its local QA/RA personnel/customers in the affected countries by email on 03/11/2016. The letter described the product, problem and actions to be taken. The customers were instructed to locate and quarantine affected unused devices immediately; return quarantined product to your national Smith & Nephew agency/distributor; complete and return the Return Slip via fax to your national Smith & Nephew agency/distributor; and pass on this safety information to all those who need to be aware in your organization. The local QA/RA staff will then notify end customers according to local procedures. In Germany as well as in Switzerland, sales representatives will inform the affected customers in-person and retrieve affected devices from their stock. If you have any questions, call 978-749-1330 or email: [email protected]

Recall details

Recall number
Z-1537-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Mar 11, 2016
Posted by the FDA
Apr 21, 2016
Recalling firm
Smith & Nephew, Inc.
Quantity
2 units
Product code
JWH
FDA submissions
K072858
Distribution
International Distribution only in countries of: Germany and Switzerland.

Product description

TC-PLUS(TM), Primary Fixed Tibial Component VKS Left, SIZE 2 LEFT, STERILE R, REF 12 000 002, S&N 7500257, Smith & Nephew Orthopaedics AG Product Usage: Tibial knee prosthesis

Code information (models, lots, serial numbers)

Batch Number C1507852

Other devices under the same FDA submission

1 device cleared under K072858 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 73609).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page