Skip to content

VITROS recallZ-1914-2016 · Ortho-Clinical Diagnostics · initiated Apr 6, 2016

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Ortho Clinical Diagnostics confirmed a software timing anomaly in which the VITROS 3600 and 5600 Systems may unexpectedly aspirate from or dispense into an unintended sample that may lead to erroneous or contaminated results.

FDA root cause category: Software design

What the company is doing

On 4/6/2016,URGENT PRODUCT CORRECTION NOTIFICATION letters (Ref. CL2016-076, dated 4/06/2016) were sent via FedEx overnight and/or US Postal Service Priority Mail (only to those consignees who provided a PO Box as an address) to all customers who received the affected product and advised them of the issue and the required actions. Follow enclosed instructions to help decrease the probability of the software anomaly until the software update is installed. Discuss any concerns with previously report results with their Laboratory Medical Director to determine the appropriate course of action. Install Software Version 3.2.3 at their earliest convenience. Foreign affiliates were informed of the issue by e-mail on 4/6/2016, and instructed to notify their consignees of the issue and required actions.

Recall details

Recall number
Z-1914-2016
Status
Terminated
Initiated by the company
Apr 6, 2016
Recalling firm
Ortho-Clinical Diagnostics
Quantity
VITROS 5600: US: 1161 units, Foreign: 1063 units; VITROS 5600 Refurbished: US: 19 units, Foreign: 38 units
Product code
JJE
FDA submissions
K081543
Distribution
Nationwide, Puerto Rico and Foreign distribution to the following countries: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.

Product description

VITROS 5600 Integrated System, Catalog Number 6802413, Unique Device Identifier No. 10758750002740; and VITROS 5600 Integrated System (refurbished), Catalog Number 6802915, Unique Device Identifier No. 10758750007110; IVD.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Software Version 3.2.2 & below; VITROS 5600 Serial Numbers J56000024 - J56002387; VITROS 5600 (Refurbished) US Serial Numbers: 56000186, 56000308, 56000315, 56000327, 56000339, 56000458, 56000559, 56000718, 56000727, 56000741, 56000747, 56000750, 56000786, 56000792, 56000824, 56001062, 56001235, 56001635, 56001641; Foreign Serial Numbers: 56000147, 56000200, 56000214, 56000245,56000283, 56000330, 56000331, 56000362, 56000377, 56000399, 56000413, 56000416, 56000428, 56000437, 56000438, 56000439, 56000459, 56000463, 56000464, 56000466, 56000515, 56000528, 56000537, 56000555, 56000560, 56000632, 56000642, 56000669, 56000680, 56000749, 56000867, 56001133, 56001135, 56001246, 56001461, 56001482, 56001487, 56001859.  Unique Device Identifier No. 10758750002740

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 73887).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page