Skip to content

BioSync Anatomic Cotton Wedge, 16 x 4.5 mm recallZ-1958-2016 · Arthrex, Inc. · initiated Apr 22, 2016

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Arthrex, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for polystyrene particulate presence.

FDA root cause category: Material/Component Contamination

What the company is doing

On 04/22/2016 Arthrex contacted all distributors that received potentially affected devices and instructed them to place the product in quarantine. Distributors were then instructed to return affected devices to Arthrex. Surgeons who have implanted the device will also be receiving a recall notification. Surgeons, risk managers, and distributors received mailed recall notification identifying the recalled products and batches, the reason for recall, and instructions for returning affected product. Customers are to call Arthrex's Customer Service Line (800-934-4404) to obtain return authorization and further instructions. Customers were also required to complete a "Recall Acknowledgement of Receipt" postal card to return to Arthrex, Inc. in an enclosed envelope.

Recall details

Recall number
Z-1958-2016
Status
Terminated
Initiated by the company
Apr 22, 2016
Recalling firm
Arthrex, Inc.
Quantity
138 total products
Product code
PLF
FDA submissions
K151256
Distribution
Distributed in the states of CA, CO, FL, GA, IL, IN, KS, MD, NC, NY, PA, SC, VA, WA, and WI.

Product description

BioSync Anatomic Cotton Wedge, 16 x 4.5 mm.; Used for internal bone fixation for bone fractures, fusions or osteotomies in the ankle and foot.

Code information (models, lots, serial numbers)

Catalog #: AR-8948W-1645; Batch #: 10019815 & 10031866; Unique Device Identifier: 00888867202702.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 74084).