Skip to content

TIGR Matrix Surgical Mesh, Model No. NSTM1015, NSTM1520… recallZ-2603-2016 · Novus Scientific Ab · initiated Jul 19, 2016

Class II Terminated All recalls posted in August 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Novus Scientific Ab or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Novus Scientific is implementing additions to the product's instructions for use concerning certain inguinal hernia repairs.

FDA root cause category: Other

What the company is doing

An Urgent Safety Alert letter dated 7/19/16 was sent to all their customers to inform them that Novus Scientific is implementing additions to the product's instructions for use (IFU) with a new contraindication and warning related to the clinical study result. The letter informs the customers that until the new IFU is made available, surgeons using the product for hernia repair should follow the instructions in the urgent safety alert in addition to those in the current IFU. Customers are informed of the actions to be taken: " Additional contraindication: Not suitable for the repair of direct inguinal hernias. " Additional warning: Because TIGR Matrix Surgical Mesh is fully resorbable, it should not be used in repairs where permanent support from the mesh is required. Customers with any questions or concerns are instructed to contact Novus Scientific AB ( +46 (0)18 700 11 50.

Recall details

Recall number
Z-2603-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 19, 2016
Posted by the FDA
Aug 19, 2016
Recalling firm
Novus Scientific Ab
Quantity
3458 units total (657 units in US)
Product code
FTL
FDA submissions
K092224
Distribution
US : IL, CA, FL, KY, MO, NH, VA, TX, NC, OH, NY, TN, Worldwide: Sweden, Italy, Denmark, UK, Singapore, Germany, Belgium, Spain, Austria, Israel, Malta, Switzerland, Ireland, Norway, Netherlands, Turkey, New Zealand, Bulgaria, France, Iran, Finland, Poland, South Africa

Product description

TIGR Matrix Surgical Mesh, Model No. NSTM1015, NSTM1520, NSTM2030

Code information (models, lots, serial numbers)

n/a

Other devices under the same FDA submission

1 device cleared under K092224 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page