Skip to content

SOMATOM Definition Flash with software version VA48A-SP2 recallZ-2625-2016 · Siemens Medical Solutions USA, Inc · initiated Jul 13, 2016

Class II Terminated All recalls posted in August 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to an internal communication error between the firmware and the software of the components, the planned CT scan executes properly, but the injector is not started. Therefore, the contrast agent is not injected and the desired examination result is not achieved. This error only affects the automatic mode or coupled mode and does not affect the manual control of the injector.

FDA root cause category: Software in the Use Environment

What the company is doing

Siemens mailed to customers an Important Customer Safety Notice to inform about a possible malfunction when using the affected systems and what actions to take to eliminate the problem. Also, Siemens informed their customers that they were currently developing a new firmware to eliminate the internal communication error, which should be released in August 2016. Customers were asked to contact their service organization at 800-888-7436, if they had any questions.

Recall details

Recall number
Z-2625-2016
Recall class
Class II
Status
Terminated
Initiated by the company
Jul 13, 2016
Posted by the FDA
Aug 23, 2016
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
2 systems
Product code
JAK
FDA submissions
K143416
Distribution
Distributed to: MI, NY, CA, KY, ND, NE

Product description

SOMATOM Definition Flash with software version VA48A-SP2; Model # 10590000, computed tomography x-ray system

Code information (models, lots, serial numbers)

Model # 10590000 Serial # 83398, and 83387

Other devices under the same FDA submission

1 device cleared under K143416 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 74840).