Xhibit Central Station, Model 96102 recallZ-0332-2017 · Spacelabs Healthcare Inc · initiated Oct 18, 2016
Class II Terminated All recalls posted in November 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The firm received reports of telemetry SpO2 numerics dropping off the Xhibit Central display. Desaturation, high, and low limit alarms work normally.
FDA root cause category: Software design
What the company is doing
The firm sent the Urgent - Medical Device Correction letter, dated 18 October 2016, via priority service, return receipt requested, to all U.S. customers on 10/18/16. The firm will e-mail a customer letter (translated as necessary) to all international subsidiaries and distributors of record on October 25, 2016. Xhibit Central Stations used without Xhibit Telemetry Receivers, are not affected. Customers were asked to circulate the letter to all person affected and place a copy with the appropriate monitor user manual until corrections have been implemented. Spacelabs Healthcare will contact affected consignees to schedule a convenient time for Spacelabs to update the affected Xhibit Central Station(s) with a new software version 1.1.6 at no cost. Customers with questions can call Technical Support at 1-800-522-7025 and select 2.
Recall details
- Recall number
- Z-0332-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 18, 2016
- Posted by the FDA
- Nov 2, 2016
- Recalling firm
- Spacelabs Healthcare Inc
- Quantity
- 348 units (292 units in the US and 56 units outside the US)
- Product code
- MHX
- FDA submissions
- K122146
- Distribution
- Worldwide Distribution -- USA, to AL, CA, CT, FL, GA, IA, ID, IL, IN, KY, MI, MN, MS, NC, NE, NJ, NY, OH, OK, PA, RI, SC, SD, TX, VA, WA, WY, and Puerto Rico; and, to the countries of AUSTRALIA, CANADA, CHILE, CZECH REPUBLIC, FRANCE, MEXICO, NETHERLANDS, PANAMA, SAUDI ARABIA, SWITZERLAND, TAIWAN, and UNITED KINGDOM.
Product description
Xhibit Central Station, Model 96102. Intended use is to provide clinicians with central monitoring of adult, pediatric and neonatal patient data of patients connected to networked Spacelabs Healthcare patient monitors and telemetry transmitters.
Code information (models, lots, serial numbers)
Show the full code information (5k characters)
Software version 1.1.5, Affected Serial Numbers in the US ************************************* ADDITIONAL FIVE UNITS ADDED: SLA0216F01796; SLA0216F01808; SLA1215F01638; SLA1215F01645; SLA1215F01652; SLA0114F00445, SLA0114F00446, SLA0114F00447, SLA0115F00917, SLA0115F00918, SLA0115F00919, SLA0115F00921, SLA0115F00923, SLA0115F00924, SLA0115F00927, SLA0115F00928, SLA0115F00934, SLA0115F00955, SLA0115F00961, SLA0115F00968, SLA0115F00969, SLA0115F00972, SLA0115F00976, SLA0116F01718, SLA0116F01720, SLA0116F01722, SLA0214F00489, SLA0216F01807, SLA0216F01809, SLA0216F01813, SLA0216F01821, SLA0216F01824, SLA0216F01837, SLA0216F01840, SLA0216F01844, SLA0314F00534, SLA0314F00535, SLA0314F00540, SLA0314F00543, SLA0314F00547, SLA0314F00548, SLA0314F00549, SLA0314F00552, SLA0314F00561, SLA0314F00567, SLA0414F00568, SLA0414F00585, SLA0416F01850, SLA0416F01853, SLA0416F01857, SLA0416F01860, SLA0416F01875, SLA0416F01876, SLA0416F01883, SLA0416F01885, SLA0416F01889, SLA0416F01895, SLA0416F01923, SLA0516F01969, SLA0516F01970, SLA0516F01972, SLA0516F01974, SLA0516F01975, SLA0516F01976, SLA0516F01977, SLA0516F01978, SLA0516F01981, SLA0516F01982, SLA0516F01985, SLA0516F01986, SLA0516F01987, SLA0516F01992, SLA0516F02002, SLA0516F02005, SLA0614F00631, SLA0614F00654, SLA0615F01188, SLA0615F01194, SLA0615F01197, SLA0615F01198, SLA0615F01199, SLA0615F01213, SLA0615F01237, SLA0615F01238, SLA0615F01239, SLA0615F01240, SLA0615F01244, SLA0615F01248, SLA0615F01261, SLA0714F00678, SLA0714F00688, SLA0714F00694, SLA0714F00702, SLA0714F00703, SLA0715F01274, SLA0715F01310, SLA0715F01314, SLA0715F01325, SLA0715F01331, SLA0715F01334, SLA0715F01346, SLA0715F01347, SLA0716F02009, SLA0716F02011, SLA0716F02013, SLA0716F02015, SLA0716F02016, SLA0716F02017, SLA0716F02020, SLA0716F02021, SLA0716F02022, SLA0716F02028, SLA0716F02034, SLA0716F02042, SLA0716F02043, SLA0716F02044, SLA0716F02045, SLA0814F00713, SLA0815F01371, SLA0815F01372, SLA0815F01373, SLA0815F01374, SLA0815F01375, SLA0815F01376, SLA0815F01379, SLA0815F01380, SLA0815F01381, SLA0815F01382, SLA0815F01383, SLA0815F01384, SLA0815F01386, SLA0815F01387, SLA0815F01388, SLA0815F01390, SLA0815F01391, SLA0815F01395, SLA0815F01397, SLA0815F01400, SLA0815F01402, SLA0815F01403, SLA0815F01404, SLA0815F01405, SLA0815F01406, SLA0815F01407, SLA0815F01408, SLA0815F01409, SLA0815F01412, SLA0815F01415, SLA0815F01418, SLA0815F01419, SLA0815F01424, SLA0816F02048, SLA0816F02050, SLA0816F02051, SLA0816F02052, SLA0816F02053, SLA0816F02054, SLA0816F02055, SLA0816F02056, SLA0816F02057, SLA0816F02058, SLA0816F02059, SLA0816F02060, SLA0816F02061, SLA0816F02062, SLA0816F02063, SLA0816F02065, SLA0816F02066, SLA0816F02067, SLA0816F02074, SLA0816F02075, SLA0816F02076, SLA0816F02077, SLA0816F02078, SLA0816F02081, SLA0816F02082, SLA0816F02083, SLA0816F02085, SLA0816F02087, SLA0816F02089, SLA0816F02091, SLA0816F02092, SLA0816F02093, SLA0816F02094, SLA0816F02095, SLA0816F02096, SLA0816F02097, SLA0816F02098, SLA0816F02099, SLA0816F02100, SLA0816F02101, SLA0816F02102, SLA0816F02103, SLA0816F02104, SLA0816F02105, SLA0816F02106, SLA0816F02107, SLA0915F01439, SLA0915F01442, SLA0915F01447, SLA0915F01451, SLA0915F01452, SLA0915F01457, SLA0915F01459, SLA0915F01462, SLA0915F01468, SLA0915F01474, SLA0915F01493, SLA0915F01496, SLA0915F01497, SLA0915F01498, SLA0915F01499, SLA0915F01501, SLA0915F01503, SLA0915F01506, SLA0915F01507, SLA0915F01508, SLA0915F01510, SLA0915F01511, SLA0915F01512, SLA0915F01513, SLA0915F01514, SLA0916F02108, SLA0916F02110, SLA0916F02111, SLA0916F02112, SLA0916F02114, SLA0916F02115, SLA0916F02117, SLA0916F02121, SLA0916F02122, SLA0916F02123, SLA0916F02126, SLA0916F02127, SLA0916F02130, SLA0916F02131, SLA0916F02132, SLA0916F02133, SLA0916F02134, SLA0916F02136, SLA0916F02137, SLA0916F02138, SLA0916F02139, SLA0916F02141, SLA0916F02142, SLA0916F02143, SLA0916F02145, SLA0916F02146, SLA0916F02147, SLA0916F02148, SLA0916F02149, SLA0916F02150, SLA0916F02153, SLA0916F02154, SLA0916F02155, SLA0916F02156, SLA0916F02158, SLA0916F02159, SLA0916F02160, SLA0916F02162, SLA0916F02163, SLA0916F02165, SLA0916F02166, SLA1015F01525, SLA1015F01528, SLA1015F01530, SLA1015F01540, SLA1015F01551, SLA1015F01560, SLA1114F00795, SLA1114F00820, SLA1115F01578, SLA1115F01579, SLA1115F01582, SLA1115F01584, SLA1115F01593, SLA1115F01594, SLA1115F01595, SLA1115F01596, SLA1115F01597, SLA1115F01599, SLA1115F01602, SLA1115F01603, SLA1115F01608, SLA1115F01609, SLA1214F00858, SLA1214F00871, SLA1214F00876, SLA1214F00881, SLA1214F00885, SLA1214F00886, and SLA1215F01632. ***********Affected Serial Numbers OUTSIDE the US ************************** SLA0114F00430, SLA0114F00437, SLA0114F00438, SLA0114F00439, SLA0114F00440, SLA0114F00443, SLA0115F00975, SLA0116F01724, SLA0314F00528, SLA0314F00541, SLA0414F00596, SLA0416F01872, SLA0416F01877, SLA0416F01880, SLA0612F00235, SLA0715F01270, SLA0715F01271, SLA0715F01273, SLA0715F01283, SLA0715F01335, SLA0716F02023, SLA0716F02025, SLA0716F02029, SLA0716F02031, SLA0716F02035, SLA0716F02036, SLA0716F02037, SLA0716F02040, SLA0716F02041, SLA0716F02047, SLA0815F01431, SLA0816F02049, SLA0816F02064, SLA0816F02069, SLA0816F02070, SLA0816F02071, SLA0816F02072, SLA0816F02073, SLA0816F02079, SLA0816F02080, SLA0816F02084, SLA0816F02086, SLA0916F02109, SLA0916F02118, SLA0916F02119, SLA0916F02120, SLA0916F02128, SLA0916F02129, SLA0916F02140, SLA0916F02161, SLA1014F00776, SLA1113F00348, SLA1113F00368, SLA1115F01604, SLA1115F01605, and SLA1213F00389.
Other devices under the same FDA submission
3 devices cleared under K122146 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Xhibit Xhibit CentralSpacelabs Healthcare (Washington), Inc · System, network and communication, physiological monitors3 deaths218 reports
- Xhibit 96102Spacelabs Healthcare (Washington), Inc · MONITOR, PHYSIOLOGICAL, PATIENT(WITH ARRHYTHMIA DETECTION OR ALARMS)1 recall1 death50 reports
- Xhibit Exhibit CentralSpacelabs Healthcare (Washington), Inc · SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS4 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page