Mevion S250 recallZ-0411-2017 · Mevion Medical Systems, Inc. · initiated Oct 20, 2016

Class II Terminated All recalls posted in November 2016

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software defect:2D projection of contours in Verity. The defect causes structures to appear off from their true position in the DRR and the radiograph with an error that increases with the distance of the structure from isocenter

FDA root cause category: Software design

What the company is doing

Mevion issued an Important Safety Notice letter on October 14, 2016 to their customers. The letter identified the affected product, problem, containment, mitigation and actions taken. Customers were informed that Mevion will be updating software when it is released. For further Information, contact: [email protected].

Recall details

Recall number
Z-0411-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Oct 20, 2016
Posted by the FDA
Nov 15, 2016
Recalling firm
Mevion Medical Systems, Inc.
Quantity
6 units
Product code
MUJ
FDA submissions
K120676
Distribution
US Nationwide Distribution in the states of: FL, MO, NJ, OH, OK

Product description

MEVION S250 Product Usage: Proton Radiation Therapy System

Code information (models, lots, serial numbers)

Serial Numbers: S250-0001 through S250-0006