Mevion S250 recallZ-0409-2017 · Mevion Medical Systems, Inc. · initiated Oct 31, 2016
Class II Terminated All recalls posted in November 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Completing Setup Without Program Move Complete Can Cause An Isocenter Error
FDA root cause category: Software design
What the company is doing
Mevion issued an Important Safety Notice letter on October 31, 2016 to customers. The identified the affected product, problem, mitigation and advisory information. The letter informed customers that Mevion will release and install a software patch that removes the reliance on tracking isocenter changes for the correct reporting of the Couch position to the user displays and to the OIS. For further information mailto:[email protected]
Recall details
- Recall number
- Z-0409-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 31, 2016
- Posted by the FDA
- Nov 15, 2016
- Recalling firm
- Mevion Medical Systems, Inc.
- Quantity
- 6 units
- Product code
- MUJ
- FDA submissions
- K120676
- Distribution
- US Nationwide in the states of FL, MO, NJ and OK
Product description
MEVION S250 Product Usage: Proton Radiation Therapy
Code information (models, lots, serial numbers)
Serial NUmbers: S250-0001 through S250-0006
Other devices under the same FDA submission
1 device cleared under K120676 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Mevion Medical Systems S250Mevion Medical Systems, Inc. · System, Radiation Therapy, Charged-Particle, Medical4 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page