CATH LAB KIT W/3 PORT "ON" MANIFOLD, WASTE BAG, SYRINGE &… recallZ-0574-2017 · ICU Medical, Inc. · initiated Sep 27, 2016
Class II Terminated All recalls posted in November 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ICU Medical Inc. has identified a potential for leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers.
FDA root cause category: Device Design
What the company is doing
ICU Medical sent an Urgent Medical Device Recall Notification letter dated September 27, 2016, to all affected customers to inform them of the expanded recall of the monitoring kits with Transpac IV transducers. Customers were informed of updated communications from August 10, 2016, when ICU Medical sent a communication indicating a potential leak with certain SafeSet Reservoir and blood sampling kits containing the Transpac IV transducers. Based on continued evaluation, ICU Medical is expanding the notification to now include certain other monitoring kits that contain the Transpac IV transducers. A small number of these devices may have the potential to leak or potentially introduce small amounts of air into the line. Customers with questions were instructed to contact ICU Customer Service Monday through Friday between the hours of 8:30 AM and 4:00 PM Pacific time: 866-829-9025 and select option 2, or e-mail the following address: [email protected]. Customers are instructed to complete the attached Medical Device Recall Response Form and fax to (801) 264-1755 or e-mail to [email protected]. Customers with questions were instructed to call 866-829-9025.
Recall details
- Recall number
- Z-0574-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Sep 27, 2016
- Posted by the FDA
- Nov 15, 2016
- Recalling firm
- ICU Medical, Inc.
- Quantity
- 30 units
- Product code
- DRS
- FDA submissions
- K061573
- Distribution
- Nationwide Distribution
Product description
CATH LAB KIT W/3 PORT "ON" MANIFOLD, WASTE BAG, SYRINGE & TP4, Item No. AG7891-NS The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a catheter and pressure monitor by converting changes in pressure into electrical currents that can be input into a compatible pressure monitor. The SafeSet Reservoir system is a blood sampling system assembled directly to the patient monitoring line (Transpac units). The SafeSet reservoir is used on patients where periodic blood sampling is required.
Code information (models, lots, serial numbers)
3280366
Other devices under the same FDA submission
22 devices cleared under K061573 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ICU Medical 46104-07Icu Medical, Inc. · Transducer, blood-pressure, extravascular7 reports
- ICU Medical 42650-06Icu Medical, Inc. · Transducer, blood-pressure, extravascular5 reports
- ICU Medical 42648-06Icu Medical, Inc. · Transducer, blood-pressure, extravascular4 reports
- ICU Medical 42598-01Icu Medical, Inc. · Transducer, blood-pressure, extravascular3 reports
- ICU Medical 46115-65Icu Medical, Inc. · Transducer, blood-pressure, extravascular3 reports
- ICU Medical 42585-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 42632-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 42590-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 852-011-46104-58Icu Medical, Inc. · Transducer, blood-pressure, extravascular2 reports
- ICU Medical 42408-11Icu Medical, Inc. · Transducer, blood-pressure, extravascular1 report
- ICU Medical 42586-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular1 report
- ICU Medical 42582-05Icu Medical, Inc. · Transducer, blood-pressure, extravascular1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 75491).
- Z-0414-2017 · Terminated4 Port "ON" Manifold (600 psi) w/4 ft. Transpac IV, Item No. 46300-14NS The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that int…
- Z-0415-2017 · TerminatedTRANSPAC IV MONITORING KIT, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP, Item No. 011-42584-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascu…
- Z-0416-2017 · TerminatedTranspac IV Bifurcated Monitoring Kit 48", 2 Disposable Transducers, 2 03 ml Intraflo Flush Devices, Macrodrip (Pole Mount), Item No. 011-42592-05 The Transpac Disposable Straight …
- Z-0417-2017 · TerminatedTRANSPAC IV MONITORING KIT, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP, Item No. 011-46106-45 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascu…
- Z-0418-2017 · TerminatedTRANSPAC IV MONITORING KIT WITH 2 BONDED STOPCOCKS, 72", DISPOSABLE TRANSDUCER WITH 2 BONDED STOPCOCKS, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT), Item No. 42500-02 The Transpac D…
- Z-0419-2017 · TerminatedTRANSPAC IV DISPOSABLE TRANSDUCER WITH STOPCOCKS, Item No. 42500-03 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfa…
- Z-0420-2017 · TerminatedTRANSPAC IV BIFURCATED KIT WITH SAFESET" RESERVOIR AND BLOOD SAMPLING PORTS, 84" TUBING, 2 SQUEEZE FLUSHES, 2 DISPOSABLE TRANSDUCER, MACRODRIP (POLE MOUNT), Item No. 42500-06 The T…
- Z-0421-2017 · TerminatedTRANSPAC IV BIFURCATED MONITORING KIT, 72" 2 DISPOSABLE TRANSDUCERS, with 2 STOPCOCKS, 2 3 ml SQUEEZE FLUSH, MACRODRIP (POLE MOUNT), Item No. 42500-19 The Transpac Disposable Strai…
- Z-0422-2017 · TerminatedTRANSPAC IV DISPOSABLE TRANSDUCER, Item No. 42582-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfaces between a c…
- Z-0423-2017 · TerminatedTRANSPAC IV DISPOSABLE TRANSDUCER WITH STOPCOCKS, Item No. 42582-06 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transducer that interfa…
- Z-0424-2017 · TerminatedTRANSPAC IV DISPOSABLE TRANSDUCER WITH 4 FT. CABLE EXTENSION AND STOPCOCK, Item No. 42582-08 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressur…
- Z-0425-2017 · TerminatedTRANSPAC IV DISPOSABLE TRANSDUCER MONITORING KIT WITH BONDED STOPCOCK AND 3mL SQUEEZE FLUSH, Item No. 42582-09 The Transpac Disposable Straight Pressure Transducer (DSPT) is an ex…
- Z-0426-2017 · TerminatedTRANSPAC IV MONITORING KIT DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (INTENDED FOR PATIENT MOUNT), Item No. 42583-05 The Transpac Disposable Straight Pressure Transducer…
- Z-0427-2017 · TerminatedTranspac IV Monitoring Kit, 60", Disposable Transducer, 3 ml Squeeze Flush Device, Macrodrip, Item No. 42584-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is an ex…
- Z-0428-2017 · TerminatedTRANSPAC IV MONITORING KIT DISPOSABLE TRANSDUCER WITH BONDED STOPCOCK, 3 ML SQUEEZE FLUSH, MACRODRIP, 60", DISPOSABLE TRANSDUCER WITH BONDED STOPCOCK, 3 ML SQUEEZE FLUSH, MACRODRIP…
- Z-0429-2017 · TerminatedTranspac IV Monitoring Kit, 84", Disposable Transducer, 3 ml Squeeze Flush, Macrodrip, Item No. 42585-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravasc…
- Z-0430-2017 · TerminatedTRANSPAC IV MONITORING KIT NEONATAL, 24", DISPOSABLE TRANSDUCER, 30 ML SQUEEZE FLUSH (INFUSION PUMP), Item No. 42586-05 The Transpac Disposable Straight Pressure Transducer (DSPT) …
- Z-0431-2017 · TerminatedTRANSPAC IV MONITORING KIT, 60", DISPOSABLE TRANSDUCER, 3 ML INTRAFLO FLUSH, MACRODRIP (POLE MOUNT), Item No. 42587-05 The Transpac Disposable Straight Pressure Transducer (DSPT) i…
- Z-0432-2017 · TerminatedTRANSPAC IV MONITORING KIT, NEONATAL, 24", WITH DISPOSABLE TRANSDUCER, 30 ML INTRAFLO FLUSH (INFUSION PUMP), Item No. 42588-05 The Transpac Disposable Straight Pressure Transducer …
- Z-0433-2017 · TerminatedTRANSPAC IV MONITORING KIT, 9", DISPOSABLE TRANSDUCER, 3 ML INTRAFLO FLUSH, MACRODRIP (PATIENT MOUNT), Item No. 42589-05 The Transpac Disposable Straight Pressure Transducer (DSPT)…
- Z-0434-2017 · TerminatedTRANSPAC IV TRIFURCATED MONITORING KIT 60 INCH, 3 DISPOSABLE TRANSDUCERS, 3 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT), Item No. 42590-05 The Transpac Disposable Straight Pressure …
- Z-0435-2017 · TerminatedTRANSPAC IV TRIFURCATED MONITORING KIT, 84" 3 DISPOSABLE TRANSDUCERS WITH 2 STOPCOCKS 3 3 ML SQUEEZE FLUSHES, MACRODRIP (POLE MOUNT), Item No. 42590-08 The Transpac Disposable Stra…
- Z-0436-2017 · TerminatedTranspac IV Monitoring Kit w/4 Way Stopcock, 3 Way Stopcock and 24" Arterial Pressure Tubing, Item No. 42594-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is an ex…
- Z-0437-2017 · TerminatedTranspac IV Monitoring Kit w/4 Way Stopcock, 3 Way Stopcock and 24" Arterial Pressure Tubing, Item No. 42594-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is an ex…
- Z-0438-2017 · TerminatedMonitoring Kit w/TP4, 30 ml Squeeze Flush and Needleless Valve, Item No. 42598-01 The Transpac Disposable Straight Pressure Transducer (DSPT) is an extravascular pressure transduce…
- Z-0439-2017 · TerminatedTRANSPAC IV MONITORING KIT WITH DISPOSABLE TRANSDUCER, 3 ML INTRAFLO FLUSH AND MACRODRIP (POLE MOUNT), Item No. 42606-05 The Transpac Disposable Straight Pressure Transducer (DSPT)…
- Z-0440-2017 · TerminatedTRANSPAC IV MONITORING KIT, 60", DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MICRODRIP (POLE MOUNT), Item No. 42607-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is…
- Z-0441-2017 · TerminatedTRANSPAC IV MONITORING KIT , 60", DISPOSABLE TRANSDUCER, 3 ML INTRAFLO FLUSH, MICRODRIP (POLE MOUNT), Item No. 42608-05 The Transpac Disposable Straight Pressure Transducer (DSPT) …
- Z-0442-2017 · TerminatedTRANSPAC IV MONITORING KIT WITH 2 SAFESET" BLOOD SAMPLING PORTS, 60" TUBING, DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT), Item No. 42620-08 The Transpac Dispo…
- Z-0443-2017 · TerminatedTRANSPAC IV MONITORING KIT, 72", DISPOSABLE TRANSDUCER, 3 ML SQUEEZE FLUSH, MACRODRIP (POLE MOUNT), Item No. 42632-05 The Transpac Disposable Straight Pressure Transducer (DSPT) is…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page