LMA MADgic, Atomization Device with 3mL Syringe, Rx only recallZ-0384-2017 · Teleflex Medical · initiated Oct 27, 2016
Class II Terminated All recalls posted in November 2016
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Teleflex Medical or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication.
FDA root cause category: Process control
What the company is doing
Teleflex sent an Urgent Medical Device Recall Notification letter on/about October 27, 2016. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and quarantine affected products. Customers were asked to complete the enclosed Recall Acknowledgement Form and fax it to Teleflex Customer Service or email to [email protected]. For questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990. Teleflex sent an Amended Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to continue to return affected product for a full refund per the procedure set forth in the letter of October 27, 2016. Some customers have indicated that, due to medical necessity, they plan to continue using the affected lots rather than return them. Teleflex is advising all such customers to follow the supplemental instructions for Use included as Attachment B. These supplemental instructions allow non-destructive testing of each unit prior to the procedure to determine if it is defective. Institutions that intend to continue using affected lots with this supplemental testing, please send the acknowledgment form to [email protected] with their name and contact information and the lots that they will continue to use, or call 866-246-6990. Customers with questions should contact their local sales representative or Customer Service at 1-866-246-6990.
Recall details
- Recall number
- Z-0384-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Oct 27, 2016
- Posted by the FDA
- Nov 9, 2016
- Recalling firm
- Teleflex Medical
- Quantity
- 2,674,650 units in total
- Product code
- CCT
- FDA submissions
- K153470
- Distribution
- Worldwide distribution. U.S. (Nationwide) and Internationally to Canada, Belgium, Singapore and Thailand.
Product description
LMA MADgic, Atomization Device with 3mL Syringe, Rx only, Product Usage: These products are used for the delivery of topical anesthesia via an atomized spray to the oral, nasal, pharyngeal or laryngeal mucosa. Teleflex Medical is recalling these products as they may produce a straight stream instead of a fully atomized plume of medication. It is unlikely that serious adverse health consequences will occur in the event of a failure to deliver an atomized plume; however, this may result in inadequate topical anesthesia which may lead to some discomfort, further attempts to deliver topical anesthesia, or the use of alternative methods of anesthesia.
Code information (models, lots, serial numbers)
Product Code: MAD600, Lot# 160110, 160119m, 160128, 160140, 160207, 160228, 160304, 160411, 160442, 160525, 160703 and 160807; Product Code: MAD700, Lot# 160111, 160129, 160141, 160209, 160233, 160316, 160329, 160403, 160431, 160502, 160520, 160604, 160624, 160634, 160712, 160809 and 160818; Product Code: MAD720, Lot# 160142, 160404, 160511, 160725 and 160909; Product Code: MAD730, Lot# 160427; Product Code: MAD730OS, Lot# 160305.
Other devices under the same FDA submission
2 devices cleared under K153470 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- LMA ATOMIZATION IPN915310Teleflex Incorporated · APPLICATOR (LARYNGO-TRACHEAL), TOPICAL ANESTHESIA2 reports
- LMA ATOMIZATION IPN048838Teleflex Incorporated · APPLICATOR (LARYNGO-TRACHEAL), TOPICAL ANESTHESIA1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 75554).
- Z-0383-2017 · TerminatedFLEXI NOZZLE SHORT Product Usage: These products are used for the delivery of topical anesthesia via an atomized spray to the oral, nasal, pharyngeal or laryngeal mucosa. Teleflex…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page