Skip to content

MEVION S250-Proton Radiation Therapy recallZ-1122-2017 · Mevion Medical Systems, Inc. · initiated Dec 16, 2016

Class II Terminated All recalls posted in February 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Mevion Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An error can occur causing Delta corrections to be lost when one setup field is closed and another is opened

FDA root cause category: Software design

What the company is doing

Mevion sent an Important Safety Notice to customers on 12/16/16 by email and Salesforce. The letter identified the affected product, problem and actions to be taken. Mevion provided mitigation and will release and install a software patch that removes the possibility that Deltas can be lost between multiple Setup fields. For further information contact [email protected].

Recall details

Recall number
Z-1122-2017
Recall class
Class II
Status
Terminated
Initiated by the company
Dec 16, 2016
Posted by the FDA
Feb 2, 2017
Recalling firm
Mevion Medical Systems, Inc.
Quantity
2 units
Product code
LHN
FDA submissions
K120676
Distribution
US Nationwide in the states of OK, NJ

Product description

MEVION S250-Proton Radiation Therapy Product Usage: Proton Radiation Therapy

Code information (models, lots, serial numbers)

S250-0002 and S250-0003

Other devices under the same FDA submission

1 device cleared under K120676 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page