LIMA Modular Revision Hip Stem Model 428-01-050_110 recallZ-1191-2017 · Encore Medical, Lp · initiated Jan 10, 2017
Class II Terminated All recalls posted in February 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Encore Medical, Lp or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Lima Proximal Bodies were inadvertently re-sterilized. The safety screw which affixes the distal and proximal bodies of the stem includes a thread-locking plug made from UHMWPE, which is not approved for repeated gamma sterilization.
FDA root cause category: Nonconforming Material/Component
What the company is doing
DJO Global sent an Urgent Field Safety Notice letter dated January 10, 2017 to customers. The letter identified the affected product, problem and actions to be taken. Customers were asked to complete the Acknowledgement and receipt Form. Customers were instructed to contact Customer Service at 1-800-456-8696 to explore options for a replacement order.
Recall details
- Recall number
- Z-1191-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 10, 2017
- Posted by the FDA
- Feb 11, 2017
- Recalling firm
- Encore Medical, Lp
- Quantity
- 13 devices
- Product code
- KWA
- FDA submissions
- K092331
- Distribution
- US Nationwide - US Nationwide in the states of MS, OH, RI, NY, OK, TX, CA, HI
Product description
LIMA Modular Revision Hip Stem Model 428-01-050_110 Product Usage: The Modular Revision Femoral hip stem is made up of a modular stem coupled with a proper neck by means of a Morse taper stabilized during the implantation phase by a safety screw. This system is particularly indicated for revision surgery on both uncemented and cemented femoral implants, when there is significant bone loss and an abnormal meta-epiphyseal anatomy of the femur.
Code information (models, lots, serial numbers)
1005528A, 1006882A, 1107305A, 1101798A, 1100565A, 1005531A, 1211716A, 1007408A, 1004038B, 1005533D, 1106805A, 1203471A, 1005533C, 1100565A, 1101795A, 1007079A, 1005532A, 1100560A
Other devices under the same FDA submission
11 devices cleared under K092331 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- DJO SURGICAL 428-16-200Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)3 reports
- DJO SURGICAL 428-01-090Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)2 reports
- DJO SURGICAL 428-01-060Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-01-080Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-20-140Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-00-080Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-00-090Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-00-100Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-01-100Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-14-200Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
- DJO SURGICAL 428-18-200Encore Medical, L.P. · Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page