6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone… recallZ-1199-2017 · Acumed LLC · initiated Jan 19, 2017
Class II Terminated All recalls posted in February 2017
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Acumed LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed.
FDA root cause category: Mixed-up of materials/components
What the company is doing
Acumed sent an Urgent Medical Device Recall letter dated January 16, 2017, to all affected customers on January 19, 2017. Consignees were instructed to follow special instructions and send survey to [email protected]. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email [email protected] to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at [email protected] or via phone 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada) have to obtain a Return Material Authorization (RMA) number.
Recall details
- Recall number
- Z-1199-2017
- Recall class
- Class II
- Status
- Terminated
- Initiated by the company
- Jan 19, 2017
- Posted by the FDA
- Feb 14, 2017
- Recalling firm
- Acumed LLC
- Quantity
- 15 units
- Product code
- HRS
- FDA submissions
- K122538
- Distribution
- Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain
Product description
6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured where beneficial to the application, and they are secured to the bone with 3.5mm non-locking screws. The Acumed pelvic ring plate is the Pubic Symphysis Plate. The Acumed Pelvic Plating System is intended for use by surgeons with orthopedic training and knowledge of the indications and techniques required for fixation. The device is to be implanted by the surgeon in a sterile operating room setting.
Code information (models, lots, serial numbers)
Lot number: 383813
Other devices under the same FDA submission
2 devices cleared under K122538 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ACUMED 30-0348Acumed LLC · Screw, fixation, bone2 reports
- ACUMED 30-0345-SAcumed LLC · Screw, fixation, bone1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page