Skip to content

The Sarns Centrifugal Pump is a single use device… recallZ-1235-2017 · Terumo Cardiovascular Systems Corporation · initiated May 27, 2015

Class II Terminated All recalls posted in February 2017

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Terumo Cardiovascular Systems Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.

FDA root cause category: Process control

What the company is doing

Customer Notification Letter/Response Form is being sent to all affected consignees.

Recall details

Recall number
Z-1235-2017
Recall class
Class II
Status
Terminated
Initiated by the company
May 27, 2015
Posted by the FDA
Feb 22, 2017
Recalling firm
Terumo Cardiovascular Systems Corporation
Quantity
579 units have entered into the distribution chain
Product code
KFM
FDA submissions
K112229
Distribution
Terumo CVS has identified 33 domestic consignees and 0 international consignee that are affected by this recall

Product description

The Sarns Centrifugal Pump is a single use device sterilized by ethylene oxide gas. The device is sold single sterile in cartons of 8 units, or may be provided in bulk, non-sterile to another Terumo location for inclusion in Cardiovascular Procedure Kits. Cardiovascular Procedure kits are packaged individually.

Code information (models, lots, serial numbers)

TAOS, TC03

Other devices under the same FDA submission

4 devices cleared under K112229 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page